日立7600与7180生化分析仪检测游离脂肪酸结果的一致性分析  被引量:1

Analysis of the Consistency of the Detection Results of Non-Esterified Fatty Acid(NEFA) by Hitachi 7600 and 7180 Biochemical Analyzers

在线阅读下载全文

作  者:吕建晓[1] 马红雨[1] 安映红[1] 罗丹[1] 朱美财[1] 

机构地区:[1]北京空军总医院临床检验中心,北京100142

出  处:《中国医药导刊》2011年第10期1776-1777,共2页Chinese Journal of Medicinal Guide

摘  要:目的:探讨日立7600与7180全自动生化分析仪检测游离脂肪酸(NEFA)结果的一致性。方法:应用两台生化分析仪对40份标本进行NEFA的测定,对检测数据进行相关性分析并计算系统误差,以评价两个检测系统结果的一致性。结果:两个检测系统NEFA测定结果相关性较好,相关系数r≥0.975,NEFA在不同医学水平的总误差都小于1/2厂家提供允许总误差(EA)。结论:日立7600与7180生化分析仪检测游离脂肪酸结果具有良好的一致性。Objective:The purpose of this paper is to analyse the consistency of the detection results of Non-Estenfied Fatty Acid (NEFA) by Hitachi 7600 and 7180 Biochemical Analyzers.Methods:Two Biochemical Analyzers were used to measure Non-Estcrified Fatty Acid of 40 samples.A relevance analysis between detected data from two analyzers was conducted and the systematic: errors were calculated, so that the consistency of the results between two detecting systems was evaluated.Results: There is a good correlation between the results of the NEFA detected by two detecting systems, with the correlation coefficient of r≥0.975.The total error of the NEEA in different medical levels was less than 1/2 of the allowed total error (EA) of the manufacturer.Conclusion:There is a good consistency in detection results of Non-Esterified Fatty Acid (NEFA) detected by Hitachi 7600 and 7180 Biochemical Analyzers

关 键 词:游离脂肪酸 生化分析仪 比对 一致性 

分 类 号:R446.1[医药卫生—诊断学]

 