作 者:王禹川[1] 丁燕生[2] 周菁[2] 盛琴慧[2] 蒋捷[2] 史力斌[2] 李康[2]
机构地区:[1]北京大学第一医院老年科,北京100034 [2]北京大学第一医院心内科,北京100034
出 处:《中国心脏起搏与心电生理杂志》2011年第6期503-505,共3页Chinese Journal of Cardiac Pacing and Electrophysiology
摘 要:目的比较主动与被动电极(AE与PE)植入右室心尖部(RVA)或右室间隔部(RVS)对即刻起搏参数的影响。方法回顾植入永久起搏器的患者资料,比较AE和PE植入RVA时即刻起搏参数的差异,比较AE植入RVA或RVS即刻起搏参数的差异。结果入选399例,283例右室起搏电极植入RVA,AE 155例,PE 128例。AE者的起搏阈值高于PE者,而阻抗、感知低于PE者,电极类型是影响差异的唯一因素。271例右室起搏电极为AE,植入RVA 155例,RVS 116例。两部位仅阻抗有差异,起搏部位是影响差异的唯一因素。结论 AE和PE在植入RVA时起搏阈值、感知和阻抗的差异由电极类型不同所引起,而AE植入RVA和RVS即刻阻抗的差异归咎于起搏部位的不同。Objective To compare the difference of the instant pacing parameters(IPPs) when inserted active electrode (AE) or passive electrode(PE) into right ventricular apex( RVA), and when inserted AE into RVA or right ventricular septum(RVS). Methods To review patients' data who received pacemaker implantation in our hospital. The statistics of the IPPs weIe analyzed when the AE was inserted into the RVA or RVS, as well as the AE or the PE was inserted into the RVA. Rt^ults Three hundred and ninety-nine patients were reviewed. There were 283 patients that the right ven- tricular electrode was inserted into the RVA. Among them, the AE was used in 155 patients while the PE was used in the others. The higher threshold was found in AE group, while the higher impedance and sensing were found in PE group. The type of electrode was the only impact factor. There were 271 patients with AE. The position of electrode insertion was at RVA in 155 patients and at RVS in the others. In two groups, the only significant difference of IPPs was found in imped- ance, and the position of electrode insertion was the impact factor. Conclusions When AE or PE is inserted into RVA, the differences in threshold, sensing and impedance are found between AE group and PE group which is attributed to the difference of electrode type. When AE is inserted into RVA or RVS, the impedance is the only difference which is attributed to the difference in pacing position.
关 键 词:心血管病学 主动电极 被动电极 右室心尖部 右室间隔 即刻起搏参数
分 类 号:R318.11[医药卫生—生物医学工程]
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.222