动力髋螺钉治疗老年股骨转子间骨折中导航技术的应用  被引量:1

Application of fluoro-navigation in the process of dynamic hip screw fixation for the femoral intertrochanteric fractures

在线阅读下载全文

作  者:周路纲[1] 王新[1] 王磊[1] 刘洪智[1] 孙煜杰[1] 苏昊[1] 

机构地区:[1]青岛大学医学院附属烟台毓璜顶医院创伤骨科,山东烟台264000

出  处:《创伤外科杂志》2012年第1期45-47,共3页Journal of Traumatic Surgery

摘  要:目的探讨导航技术在动力髋螺钉(DHS)治疗股骨转子间骨折中的价值。方法使用X线透视计算机辅助导航进行动力髋螺钉的置入、尖顶距的测量。结果 127例动力髋螺钉位置准确,尖顶距≤25mm,平均(18.62±0.69)mm,无一例切割并发症发生。结论计算机导航的瞬时追踪使术者能在术中监测螺钉位置及尖顶距,提高了螺钉置入的准确性,减少术后螺钉切割并发症。Objective To explore the clinical application of fluoro - navigation in the process of dynamic hip screw fixation for the femoral intertrochanteric fractures. Methods Totally 86 femoral intertroehanteric fractures were treated with dynamic hip screw under fluoro-navigation. The fluoro-navigation guidance was used for the entry of screws,showing screw position in the femoral neck and measuring tip-apex distance. Results The screw position of 86 femoral intertrochanteric fractures were satisfactory. The average tip-apex distance was less than 25mm,no complication was recorded during and after operation. Conclusion Fluoro-navigation helps to monitor the screw insertion and the tip-apex distance. This method has advantages of less X-ray exposure, accurate implant position, and less possibility of screw incision.

关 键 词:计算机导航 动力髋螺钉 股骨骨折 

分 类 号:R814.4[医药卫生—影像医学与核医学] R683.42[医药卫生—放射医学]

 