检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
作 者:李梅[1]
机构地区:[1]武警工程大学通信工程系,陕西西安710086
出 处:《核电子学与探测技术》2011年第12期1408-1410,共3页Nuclear Electronics & Detection Technology
摘 要:介绍了光子计数成像系统的结构、工作原理和工作平台。实验中,为了防止粒子反馈到输入端及得到更高的电子倍增,微通道板(MCP)的通道一般会倾斜6°~8°。然而,这种倾斜会引起图像偏移,影响图像精确度,为了获得清晰的图像,采取了一些措施纠正偏移,主要通过修正电子云质心的偏移,提高图像的质量。Single photon counting imaging system is introduced,and principle of workbench is described.The presence of MCP(Micro-channel Plate) pore bias induces image replacement in single photon counting imaging system.This is a precision study.In order to prevent ion feedback and acquire higher electron multiplication,MCP has their pore bias.However,the channel bias can shift the placement of image.To obtain better clear image,we take some steps to correct original image which is shifted.In the paper,through correcting the offset of centroid in charge cloud form MCP to improve the quality of image.
分 类 号:TN23[电子电信—物理电子学] O434.2[机械工程—光学工程]
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.177