室内空气中总挥发性有机物(TVOC)检验方法的探索  被引量:5

Exploration the Detection Method Indoor Air of Total Volatile Organic Compounds

在线阅读下载全文

作  者:傅瑛[1] 王亦军[1] 张凤艳 

机构地区:[1]青岛大学化学化工与环境学院,山东青岛266071 [2]青岛市产品质量监督检验所,山东青岛266071

出  处:《分析测试技术与仪器》2012年第1期47-56,共10页Analysis and Testing Technology and Instruments

摘  要:使用TVOC混合标准溶液对气相色谱法检测室内空气样品中总挥发有机物的方法进行了探索.改进后的色谱分析时间为26.1 min,比GB/T 18883-2002方法(52 min)缩短了25.9 min,TVOC各物质峰分离度高于99%,检测精密度相对标准偏差在1.05%至2.89%之间,回收率在95%到102%之间,检出限为0.020μg.Total Volatile Organic Compounds(TVOC) blended standard solution was used to explore the detection method of TVOC by GC and to determine the best chromatographic conditions.The improved process used only 26.1min,which is 25.9 min shorter than that of the original one(52 min) described in GB/T1 8883-2002.The separating degree of each component in the TVOC blended standard solution was greater than 99%.The RSD of precision was between 1.05% and 2.89%.the recovery was between 95% and 102%,and the detection limit was 0.020 μg.

关 键 词:气相色谱 总挥发性有机物 色谱条件改进 

分 类 号:O657.32[理学—分析化学]

 