作 者:马柯帆[1] 吴金杰[2] 王培玮[2] 杨元第[2] 段小娟[1]
机构地区:[1]成都理工大学,成都610059 [2]中国计量科学研究院,北京100013
出 处:《核电子学与探测技术》2012年第2期207-209,217,共4页Nuclear Electronics & Detection Technology
摘 要:采用胶片剂量计测量了低能X射线辐射装置距X射线光管焦斑不同距离辐射野的大小,得出在距离光管焦斑1 000 mm的参考面辐射野大小及均匀性,并与二维矩阵电离室扫描结果进行了对比。实验结果表明,在距X射线光源焦斑1 000 mm处,在保证直径为100 mm的均匀野的条件下,非均匀性不大于1.5%。在保证非均匀性不大于5%和50%条件下,均匀野大小分别为130 mm和145 mm,满足了低能X射线空气比释动能基准装置量值复现,国际比对和向下量值传递的要求。The size of reference radiation field of low - energy x - ray was measured by film in different position from the tube spot, the diameter of the radiation field and its were obtained in reference plane ( 1 000 mm from the radiation source) and were compared with the results through two - dimensional ion chamber. The results show that the diameter of the radiation field is 100 mm within 1.5% ( 1 000 mm from the radiation source). The diamerers of theradiation field are 130 nun and 145 mm within 5% and 50% . These results well meet the requirements of the value repetition of the low - energy X - ray air kerma value standard device, international intereomparison and transfer of quantity values.
分 类 号:TB9[一般工业技术—计量学]