便携式血糖仪与生化分析仪血糖检测结果的相关性研究  被引量:4

Correlation Studies of Glucose Testing Results between Portable Blood Glucose Meter and Biochemical Analyzer

在线阅读下载全文

作  者:孙虹[1] 苏敏[1] 蒋杰[1] 牛华[1] 蒋红君[1] 

机构地区:[1]云南省第一人民医院检验科暨云南省临床检验中心,昆明650032

出  处:《现代检验医学杂志》2012年第2期77-79,共3页Journal of Modern Laboratory Medicine

摘  要:目的 :通过对三台不同型号便携式血糖仪与生化分析仪血糖检测结果进行比对和偏差分析,了解血糖仪与生化仪血糖检测结果的相关性。方法参照《医疗机构便携式血糖检测仪管理和临床操作规范(试行)》中静脉血样比对试验方案,收集50例不同血耱浓度患者静脉全血样本,血糖仪测定全血葡萄糖,剩余样本立刻离心分离血浆,30min内用生化仪测定血浆葡萄糖,计算血糖仪与生化分析仪测定结果的相关系数和偏差。结果三台血糖仪批内CV在1.28%~4.87%之间;日问不精密度低浓度质控品s〈O.13mmol/L中、高浓度质控品CV在0.81%~2.14%之间,符合《规范》中精密度的要求。三台血糖仪与生化仪测定结果的相关系数r均〉0.98,回归方程斜率b〉1.0士0.05的理想范围。血糖浓度2.3~4.2mmol/L的样本,稳步、卓越和活力血糖仪只有60%,30%和80%检测结果误差在士0.83mmol/L的范围内;血糖浓度4.6~22.6mmol/L样本,有95%~97.5%检测结果误差在允许范围内。结论三台血糖仪精密度符合《规范》要求,与生化分析仪血糖测定结果有较好相关性。血糖仪的低浓度样本(血糖〈4.2mmol/L)误差不符合《规范》要求,必要时抽取静脉血生化仪复查;中、高浓度样本误差符合《规范》要求。Objective To investigate the glucose testing results correlation between portable blood glucose meter and bio- chemical analyzer by comparing the testing results of three different modes of portable blood glucose meters and biochemical analyzer. Methods With reference to the protocol of venous sample comparison from 《the management and clinical practices of portable blood glucose monitoring device in medical institution (Trail)》, 50 cases of heparin anti-coagulation venous whole blood were collected. Whole blood glucose was measured by glucose meter firstly,and then the remaining samples were cen- trifuged immediately to get plasma. The plasma glucose was tested by biochemical analyzer within 30 minutes. The correla- tion coefficients and deviations were calculated from the results between three glucose meters and biochemical analyzer and also the relevant analysis was conducted. Results The within-run CV of three glucose meters was between 1.28% and 4. 870/00. The intra-day imprecision of low concentration QC material was s 〈0.13 mmol/L. The CVof middle and high con- centration QC material was between 1.28 % and 4.87%, which meet the precision requirement of the specification. The re- sults coefficients of these two different testing equipments (r) were higher than 0. 98. The slope of linear regression equation b〉 1.0±0. 05 and for samples which the glucose concentration was between 2.3 and 4.2 mmol/L, the percentages of testing results within the acceptable error(±0. 83 mmol/L) were 60%, 30%and 80%for Sure Step, Performa and Active blood glucose meter, respectively. While for samples which the glucose concentration was between 4.6 and 22. 6 mmol/L,the per- centage of testing results within the acceptable error was between 95% and 97. 5%. Conclusion Results of three portable glucose meters were all correlated very well with those of biochemical analyser and meet the requirement of the specification. But results of glucose meters were lower than biochemical analyser,generally. The results de

关 键 词:便携式血糖仪 生化分析仪 血糖 偏差 相关性 

分 类 号:R446.112[医药卫生—诊断学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象