摘  要:目的制定科学合理的血液复栓规则并对其进行评价。方法以国际血液学41条复捡规则为参考,结合Beck—man-coulterAC.T5diff血液分析仪的-测试性能及报警系统,自定复检规则。对1400例门诊和住院病人的标本进行全自动血细胞分析并同时涂片镜检,观察细胞数量和形态学变化。比较二者结果,对各务自定复检规则进行统计学分析。结果各条规则中假阳性率最高的是异常淋巴细胞,其次是巨大不成熟细胞及单核细胞。真阳性率最高的是粒细胞相对计数。自定复检规则形态学评估真阳性率12.3%,假阳性率2.7%,真阴性率83.O%,假阴性率2.0%。结论制定适合Beck—man-coulter AC.T 5diff血液分析仪的15条复检规则,照此执行,既可保证工作质量又提高工作效率,为临床提供更具参考价值的信息。Objective To establish the microscope review criteria for the Beckman-coulter AC. T 5diff hematology analyzer. Methods Totally 1 400 blood samples were tested by using the Bckman-coulter AC. T 5diff hematology analyzer and smeared for microscopic examination. True positive rate, false positive rate, true negative rate and false negative rate were e- valuated and calculated according to the 41 rules suggested by the International Consensus Group for hematology review as well as the particular alarm system of the Beckman-coulter AC. T 5diff hematology analyzer. Results The most cases show- ing false positive rate were abnormal lymphocytes, large immature cells and monocytes. The most case showing true positive rate was relative granulocyte count. Microscopic examination results:true positive rate, false positive rate, true negative rate and false negative rate were respectively 12.30//00,2.70//00,83. 0% and 2.0%, respectively. Conclusion Establishment of rea- sonable microscope review criteria is beneficial to improving work efficiency, reducing probability of missed diagnosis and providing more useful information for clinical.

关 键 词:血液分析仪 复检规则 血细胞涂片 

分 类 号:R446.111[医药卫生—诊断学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象