ALISEI全自动酶免分析仪驱动电机故障的检修规律  被引量:3

Maintenance of driving motor failure of Alisei automatic ELISA analyzer

在线阅读下载全文

作  者:程广斌[1] 黄金清[1] 马承华[1] 

机构地区:[1]武警广东省总队医院医学工程科,广东广州510507

出  处:《中国医学装备》2012年第6期83-84,共2页China Medical Equipment

摘  要:目的:ALISEI全自动酶免分析仪具有开放、高速、自动化的性能和特点,将构造和使用上的规律应用于维修,可准确判断并排除故障。方法:通过案例的分析及对结构的观察,总结维修规律。结果:通过2例维修分析,ALISEI全自动酶免仪驱动电机位置较低易引发其他故障。结论:由于ALISEI全自动酶免分析仪驱动电机位置的特点,维修时应先由测量电机参数开始而后判断控制板,可以避免维修误判引发更多故障。Objective: To accurately determine and remove the faults by applying the law of the structure and use of Alisei automatic ELISA analyzer to its maintenance. Methods: Through cases analysis and the observation of the structure of Alisei automatic ELISA analyzer, the law of its maintenance was summarized. Results: Analysis of two cases of maintenance showed the driving motor of Alisei automatic ELISA analyzer was placed at a lower position, which easily led to other failures during maintenance. Conclusion: Due to the lower position of the driving motor of Alisei automatic ELISA analyzer, its maintenance should start from measuring motor parameters and then determine the control boards so as to avoid false determination leading to more maintenance problems.

关 键 词:ALISEI 全自动酶免分析仪 维修 

分 类 号:TH789[机械工程—仪器科学与技术]

 