Destiny Max全自动血凝分析仪性能评价  被引量:2

Performance Evaluation of Destiny Max Automatic Coagulation Analyzer

在线阅读下载全文

作  者:谢田刚[1] 王海燕[1] 王胜江[1] 

机构地区:[1]大连市中心医院,辽宁大连116033

出  处:《医疗卫生装备》2012年第6期107-109,共3页Chinese Medical Equipment Journal

摘  要:目的:对Destiny Max全自动血凝分析仪的机械法检测系统进行性能评价。方法:选择凝血酶原时间(PT)、部分活化凝血酶原时间(APTT)、凝血酶时间(TT)和纤维蛋白原(FIB)浓度4个指标,采用机械式磁珠法检测对准确性、精密度和携带污染率进行实验评价,并与STR-Evolution上所测结果进行相关性和Kappa一致性检验分析。结果:TriniCHECK Control 1在Destiny Max全自动血凝仪上测定的PT、APTT、TT、FIB的准确性偏差分别为2.21%、2.57%、4.68%、4.72%;TriniCHECK Control 2测定的PT、APTT、FIB准确性偏差分别为1.96%、0.54%、3.46%,精密度分别为1.33%、1.90%、3.54%;FIB高值携带污染率为2.56%,FIB对PT试剂携带污染率为2.15%;Destiny Max与STR-Evo-lution全自动血凝分析仪的临床标本比对结果良好,线性相关系数均大于0.95,K值均大于0.8,P值大于0.05,无统计学差异。结论:Destiny Max全自动血凝仪的准确性、精密度以及携带污染率都达到了性能要求,并且与STR-Evolution的临床标本比对结果有很好的相关性和一致性,可为临床提供可靠的检测数据。Objective To evaluate the detection system performance of mechanical method of Destiny Max automatic coagulation analyzer.Methods Mechanical method was selected,and prothrombin time(PT),activated partial thromboplastin time(APTT),thrombin time and fibrinogen(FIB) were used to evaluate the accuracy,precision and carry over rate.The sample test results were compared with STR-Evolution,and then the correlation and Kappa consistency analysis were evaluated.Results The accuracy deviation of PT,APTT,TT and FIB tested with TriniCHECK Control 1 on Destiny Max were 2.21%,2.57%,4.68% and 4.72%.The accuracy deviation of PT,APTT and FIB tested with TriniCHECK Control 2 were 1.96%,0.54% and 3.46%,and the precision were 1.33%,1.90% and 3.54%,respectively.The carry over rate of FIB high value was 2.56%,and that of FIB to PT reagent was 2.15%.The clinical sample results comparison between Destiny Max and STR-Evolution was favorable,the linear correlation coefficient r0.95,K0.8,P0.05,and with no significant difference.Conclusion The accuracy,precision and carry over rate of Destiny Max all have reached the performance requirements,and the clinical results compared to STR-Evolution have good relevance and consistency,so Destiny Max automatic coagulation analyzer could provide reliable test results for clinic.[Chinese Medical Equipment Journal,2012,33(6):107-109]

关 键 词:全自动血凝分析仪 准确性 精密度 携带污染率 

分 类 号:R446.1[医药卫生—诊断学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象