作 者:庞晋山[1] 黄刚[2] 邓爱华[1] 彭晓俊[1]
机构地区:[1]新会出入境检验检疫局,江门529100 [2]惠州出入境检验检疫局,惠州510020
出 处:《理化检验(化学分册)》2012年第6期671-673,677,共4页Physical Testing and Chemical Analysis(Part B:Chemical Analysis)
摘 要:选取铝、铜、镁、钙、铁、锰和锌等7种元素的9条谱线,研究电感耦合等离子体原子发射光谱法对其测定时易电离元素钾、钠的基体干扰。结果表明:易电离元素对待测元素具有较明显的干扰作用,而增大射频发射功率及降低雾化器压力可有效地抑制易电离元素对待测元素的干扰。Matrix interferences of the easily ionized elements, e. g. K and Na, with the ICP-AES determination of 7 elements (i. e., A1, Cu, Mg, Ca, Fe, Mn and Zn), were studied. Nine spectral lines of the 7 dements were taken as objectives in the study. It was {ound that the significant interferences due to the easily ionized elements (taking K and Na as example) were effectively inhibited by increasing the radio frequency transmitting power and lowering the atomizing gas pressure.
关 键 词:电感耦合等离子体原子发射光谱法 易电离元素 基体干扰