探讨数字断层融合技术层中心与图像变形的相关性  被引量:6

Assessment of correlation between layer center and image deformation in digital tomosynthesis

在线阅读下载全文

作  者:郁仁强[1] 曾勇明[1] 彭盛坤[1] 彭刚[1] 赵峰[1] 谭欢[1] 蔡文晶 

机构地区:[1]重庆医科大学附属第一医院放射科,重庆400016 [2]岛津企业管理(中国)有限公司,上海100020

出  处:《中国医学影像技术》2012年第7期1401-1405,共5页Chinese Journal of Medical Imaging Technology

基  金:重庆市卫生局科研基金(2008-2-29)

摘  要:目的通过模型评价数字断层融合(DTS)技术层中心与图像变形的相关性。方法按左右(L-R)方向和头尾(C-C)方向均相距6cm将11个直径为13.5mm钢球排在同一平面。将实验模型固定在距检查床120mm高度处,变化Height值;变化实验模型高度,同时变化对应Height值;分别测量各钢球在L-R和C-C两个方向上的直径及C-C方向上的成像范围。结果采用Thickness++重建后,各钢球L-R方向上的直径无明显变形,C-C方向上的直径均变小,最大变形度约为2.96%,位于层中心的钢球变形度最小。在C-C方向上存在金属伪影,采用Thickness++(Metal)重建后金属伪影不明显,但钢球在L-R和C-C两个方向的直径均变小。实验模型越靠近床面,C-C方向上的成像范围越大。结论 DTS偏中心成像时,ROI图像会发生变形;采用Thickness++(Metal)重建能有效去除金属伪影。Objective To observe the correlation between layer center and image deformation of digital tomosynehesis imaging by experimental model.MethodsEleven steel balls with diameter of 13.5 mm were placed in one plane,which were apart from 6 cm in the left to right(L-R) direction and cranial to caudal(C-C) direction.Height value was changed while the altitude of experimental model was fixed at 120 mm,and the altitude of experimental model and Height value were all changed,then the diameters of steel balls in L-R and C-C direction and the imaging range in the C-C direction were measured respectively.ResultsUsing "Thickness++" reconstruction,the diameters of steel balls in L-R direction had no obvious distortion,while in C-C direction decreased.The maximum deformation was about 2.96%,and the deformation of steel balls in the layer center was the minimum.There were metal artifacts in C-C direction.Using "Thickness++(Metal)" reconstruction,the metal artifacts were very weak,but the diameters of steel balls both in L-R and C-C direction decreased.The altitude of experimental model was more close to tabletop,and the imaging range in the C-C direction was larger.ConclusionAt the time of offset center imaging,deformation of ROI can happen.The metal artifacts can be removed effectively using "Thickness++(Metal)" reconstruction.

关 键 词:数字断层融合 变形度 伪影 成像质量 

分 类 号:R816[医药卫生—放射医学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象