湿度对便携式甲醛分析仪现场检测的影响  被引量:2

Humidity on portable formaldehyde analyzer the effect of field detection

在线阅读下载全文

作  者:赵海山 闻真 沈浩 

机构地区:[1]上海市环保产品质量监督检验总站

出  处:《上海计量测试》2012年第4期18-19,共2页Shanghai Measurement and Testing

摘  要:采用30m^3实验舱,在温度为(20±2)℃,对不同湿度条件下仪器现场检测的情况进行了探讨。当环境湿度在一定范围内仪器法的检测结果能符合技术要求。当湿度超过75%RH时,检测结果在某些测试浓度段出现较大偏差。当湿度超过80%RH时,检测结果已不符合技术要求。湿度对室内空气质量采用仪器法现场检测的结果有显著影响。Conducted experiments by testing the Formaldehyde Portable Detectors in an experimental room with 30m3 in space, with different levels of humidity at temperatures of (20±2)℃. Within a certain range of environment humidity, the testing results of Formaldehyde Portable Detectors met technical standards quite well. With humidity of over 75%RH, testing results deviated from the accurate value considerably. When the humidity is above 80%RH, the readings of the Formaldehyde Portable Detectors cannot fulfill technical standards. Humidity has significant impact on the accuracy of Formaldehyde Portable Detectors.

关 键 词:甲醛 湿度 现场检测 

分 类 号:X831[环境科学与工程—环境工程]

 