POCT血糖仪与全自动生化分析仪检测血葡萄糖结果对比分析  被引量:18

Comparison analysis on POCT glucose meter and automatic biochemical analyzer for detecting blood glucose

在线阅读下载全文

作  者:马骥[1] 李思逸[2] 郑智明[1] 柯培锋[1] 黄宪章[1] 

机构地区:[1]广东省中医院检验科,广州510120 [2]广东省中医院泌尿外科,广州510120

出  处:《国际检验医学杂志》2012年第14期1687-1689,共3页International Journal of Laboratory Medicine

摘  要:目的探讨POCT快速血糖仪与自动生化分析仪检测血糖结果的相关性及差异性,为临床提供参考。方法用5个浓度的质控品和患者标本在POCT血糖仪上进行检测,连续测试5d,计算变异系数(CV)。用POCT快速血糖仪和全自动生化分析仪对54例患者的末梢血和静脉血浆进行血糖检测,测得的数据用SPSS统计学软件进行分析,评价POCT血糖仪与全自动生化仪与测定结果的相关性。结果 5个浓度的质控品和患者标本在POCT血糖仪上所测结果的CV在2.6%~4.2%,低于厂家的声明,也符合国家对血糖检测系统的测量重复性要求(CV不超过7.5%)。雅培Optium Xceed血糖仪与全自动生化仪所测结果的相关系数(r)=0.974,相关性较好。结论雅培Optium Xceed血糖仪在精密度和准确度等方面均符合临床要求,其检测结果接近血浆葡萄糖检测结果。但是POCT血糖仪仅可用于患者空腹血糖的监测或血糖过筛实验,不能代替实验室血糖的检测结果作为确诊实验。Objective To discuss the difference and correlation between point of care testing(POCT) glucose meter and automatic biochemical analyzer.Methods Control materials and patients samples with 5 gradual levels were detected continually for 5 days and then the coefficient of variation(CV) was calculated.Capillary blood glucose(CBG) of 54 diabetes patients were measured by POCT glucose meter,and venous plasma glucose(VPG) of the same patients were measured as well as by hexokinase method as control.The results were measured by SPSS13.0,and the correlation between the two analyzers was evaluated.Results CV of the results of materials and patients samples with 5 gradual levels,detected by POCT,were 2.6%-4.2%,lower than the manufacturer’s statement and in accordance with the required CV below 7.5% acclaimed by states.Abbott Optium Xceed POCT glucose meter had good correlations with automatic biochemical analyzer(r=0.974).Conclusion The precision and accuracy of the POCT glucose meter were accorded with clinical requirements,and the results detected by POCT glucose meter could be closer to the plasma glucose concentration.However,POCT glucose meter could be used for monitoring the fasting blood glucose and for the blood glucose screening test.POCT could not replace the traditional glucose tests and could not be used as diagnostic test.

关 键 词:床旁检测 血糖 全自动生化分析仪 性能比较 

分 类 号:R446.1[医药卫生—诊断学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象