锁定加压接骨板四肢骨折断裂的原因分析  被引量:2

Locking compression plate fractures in the limbs of reason analysis of the fracture

在线阅读下载全文

作  者:葛金平[1] 章俊祥[1] 吴冰[1] 孙大连[1] 何成文[1] 龚有兵[1] 

机构地区:[1]安徽宣城中心医院骨科,242000

出  处:《中国实用医药》2012年第21期28-29,共2页China Practical Medicine

摘  要:目的通过观察接骨板断裂的病例,分析其断裂的原因。方法临床上共观察行锁定加压接骨板治疗的患者80例,其中,发生接骨板断裂患者6例,对该6例患者的情况进行分析,总结出发生断裂的可能原因。结果本组所有发生断裂的病例均有早期过早活动的情况,而发生断裂的原因是多方面的,有螺针使用不当、钢板选择不当、临床操作不当、解剖复位要求过渡等多方面的原因。结论正确选择螺针、钢板,规范临床操作,可以减少接骨板发生断裂的概率。Objective Through the observation of plate fracture cases, analyze the reasons for the frac- ture; Methods The clinical observation of a total of line locking compression plate treatment of 80 patients, which occurred 6 cases of plate fracture patients, the 6 patients were analyzed, summarized the possible causes of the rupture. Results The group of all fracture cases had an early premature fracture are many reasons, im- proper use of screw-pin, steel improper selection, improper operation of the clinical anatomy of the reset re- quirements of the transition, and many other reasons. Conclusion The correct choice of screw-pin, plate, and standardize the clinical practice, can reduce the probability of plate fracture.

关 键 词:锁定加压接骨板 骨折 原因 

分 类 号:R687.3[医药卫生—骨科学]

 