梯度淋洗离子色谱法测定饮用水中6种阴离子  被引量:8

Determination of Six Anions in Drinking Water by Gradient Ion Chromatography

在线阅读下载全文

作  者:戴礼洪[1] 刘岩[1] 王迪[1] 陈春[1] 瞿巧钰[1] 刘潇威[1] 

机构地区:[1]农业部环境保护科研监测所,农业部/天津市产地环境与农产品安全重点开放实验室,天津300191

出  处:《环境监测管理与技术》2012年第4期37-39,43,共4页The Administration and Technique of Environmental Monitoring

基  金:中央级公益性科研院所基本科研业务费专项基金(农业部环境保护科研监测所)资助项目(2010-WD-4)

摘  要:采用离子色谱法同时测定饮用水中F-、Cl-、Br-、NO3-、SO24-、PO34-等6种阴离子,水样经0.22μm滤膜过滤,自动淋洗发生器产生20 mmol/L~40 mmol/L KOH淋洗液梯度洗脱,在流量0.25 mL/min条件下采用IonPac AG19保护柱和IonPac AS19分离柱分离和定量。方法线性良好,6种阴离子的检出限为0.005 mg/L~0.057 mg/L,标准溶液平行测定6次,测定值的RSD在0.5%~2.2%之间,水样加标回收率在83%~106%之间。Six anions such as drinking water were simultaneously fluorinion, chloridion, bromine ion, nitrate ion, sulfate ion,phosphate ion in determined by ion chromatography. The sample water was filtered by membrane with diameter 0.22 p,m pore. Dealing with IonPac AS19 column and IonPac AG19 guard column, six anions in water were quantitative determination by ion chromatography by using 20 mmol/L - 40 mmol/L KOH generated by eluent generator at a flow rate of 0.25 mL/min. The calibration curves showed a good linearity. Detection limits of six anions ranged from 0. 005 mg/L to 0. 057 mg/L, RSD of detective values from 0.5% to 2.2% by parallel detection of six times. The spiked recoveries ranged from 83% to 106%.

关 键 词:阴离子 梯度淋洗 离子色谱法 饮用水 

分 类 号:O657.75[理学—分析化学]

 