摘  要:目的:监测分析医院船温湿度和电源电压,改善医院船上装医疗设备的储存与工作环境。方法:在执行"和谐使命-2011"任务期间,使用温湿度计、数字多用表对医院船CT室以及检验科的温度、湿度、电源电压进行监测,并进行数据统计分析。结果:温度在正常范围内,但湿度始终较高;电源电压值和电压波动范围均在正常范围内,但存在突然断电现象。结论:重要的医疗设备必须使用UPS电源,设备所在舱室应安装除湿机。Objective To monitor and analyze the temperature, humidity and voltage of the hospital ship to improve the storage and working conditions. Motbods During "Harmonious Mission 2011", thermo-hygrometer and digital muhimeter were used to monitor the temperature, humidity and voltage, then the data were analyzed statistically. Results The temperature, voltage and current were all kept in the normal ranges, while the high humidity and sudden power failure existed at the same time. Conclusion UPS power has to be available for the important medical devices, and the chamber holding medical equipment should be provided with the dehumidifier. [Chinese Medical Equipment Journal,2012,33 (9):109-110,124]

关 键 词:医院船 温湿度 电源电压 监测 分析 

分 类 号:R318.6[医药卫生—生物医学工程] R821.89[医药卫生—基础医学]

 