较低温度条件下SSE影响红外温度计校准的实验研究  被引量:2

The Experiment Study for SSE on Calibration of Infrared Thermometer under Lower Temperature Condition

在线阅读下载全文

作  者:赵泉钦[1] 李文军[1] 孙坚[1] 余时帆 章龙飞[1] 崔超[2] 

机构地区:[1]中国计量学院,浙江杭州310018 [2]浙江省计量科学研究院,浙江杭州310013

出  处:《计量学报》2012年第6期509-512,共4页Acta Metrologica Sinica

摘  要:讨论了光源尺寸效应(SSE)现象,并应用SSE系数对红外温度计的SSE特性进行估计。介绍了用于较低温度条件下测量SSE的实验装置,应用该装置对几种红外温度计进行了测试。测试结果表明,当黑体辐射源的光阑孔径大于40mm时,可以不考虑SSE的影响。该结论适用于大多数工业用温度计的校准。The phenomenon of size-of-source effect (SSE) is discussed and the coefficient of SSE is used for estimating the SSE property of the infrared thermometer. The experiment apparatus used for measuring SSE under lower temperatures condition is described, and several types of infrared thermometers are applied to the measurements of SSE. The experiment results indicate that the influence of SSE may be not considered at the diameters of the diaphragm hole being more than 40mm. This conclusion is suitable to the most industrial infrared thermometers.

关 键 词:计量学 光源尺寸效应 红外温度计 校准 黑体辐射源 SSE系数 

分 类 号:TB942[一般工业技术—计量学]

 