作 者:董磊[1] 刘娟[2] 马红雨[1] 全首祯[1] 吕建晓[1] 朱美财[1]
机构地区:[1]空军总医院临床检验中心,北京100142 [2]空军总医院输血科,北京100142
出 处:《医疗卫生装备》2012年第10期104-105,共2页Chinese Medical Equipment Journal
摘 要:目的 :对日本ARKRAY公司HA-8160型全自动糖化血红蛋白分析仪检测系统的主要分析性能进行验证。方法:对该仪器检测项目的精密度、携带污染率及线性范围进行评价。结果:检测项目各精密度在可接受范围内;携带污染率在可接受范围内;线性验证标本按一定比例稀释后所得理论值与实测值的回归方程y=bx+a中,b值介于0.97~1.03范围内,a在可接受范围内。结论:HA-8160检测系统的精密度、携带污染率及线性范围基本符合实验诊断学实验室的要求。Objective To verify the main analyzing performance of HA-8160 automatic glycosylated hemoglobin analyzer,which is produced by ARKRAY in Japan.Methods The precision,carryover rates and linearity of the selected testing items were evaluated.Results All the precision and carryover rates of the selected testing items were within an acceptable range.The regression equation between the theoretical and measured values in the linear verification experiment was y=bx+a,in which b value was within the range of 0.97~1.03,and a value was within an acceptable range.Conclusion The precision,carryover rates and linearity of the selected testing items of HA-8160 automatic glycosylated hemoglobin analyzer accord with the demands of the experiments in clinical diagnosis.
分 类 号:R318.6[医药卫生—生物医学工程] TH776[医药卫生—基础医学]