定量CT腰椎骨密度测量的低剂量研究.  被引量:10

Low-dose radiation in lumbar bone mineral density measurement by QCT

在线阅读下载全文

作  者:王予生[1] 过哲[1] 李端端[1] 赵海竹[1] 刘庆华[1] 于爱红[1] 赵英威[1] 程晓光[1] 

机构地区:[1]北京积水潭医院放射科,北京100035

出  处:《中国骨质疏松杂志》2012年第11期992-995,共4页Chinese Journal of Osteoporosis

基  金:北京市卫生系统高层次卫生技术人才培养项目(2009-02-03)

摘  要:目的通过临床验证多排螺旋定量CT腰椎低剂量扫描参数,制定最佳的低剂量多排螺旋QCT腰椎扫描协议。资料与方法选取住院拟进行腰椎手术,术前行常规多排螺旋CT平扫和腰椎QCT检查的30位患者。其中男:10例,女20例;年龄48~79岁,平均年龄63.97±8.66岁。分两次采集常规250 mA和50 mA两组腰椎加校准体模QCT容积数据,将两组L2-L4容积数据传至骨密度工作站进行骨密度(Bone Mineral Density)测量。应用统计软件SPSS 15.0进行统计分析。结果 250mA组与50 mA组L2、L3、L4平均BMD值无差异(P>0.05);两组三椎体总平均BMD值无差异(p>0.05);250 mA组辐射剂量DLP值为229.06±12.35 mGy.cm。50 mA组DLP值为45.78±2.48 mGy.cm。结论多排螺旋定量QCT腰椎低剂量扫描技术能够保证BMD测量准确性,同时使患者的受辐射剂量比常规多排螺旋CT腰椎扫描大幅度降低。Objective To investigate the influence of low dose multi-detector helical CT scan on the accuracy of lumbar spine bone mineral density (BMD) measurement by QCT. Materials and Methods Thirty adult inpatients ( 10 males and 20 females) underwent routine lumbar CT scan with 250 mA and QCT with 50 mA as standard pre-operation protocol for spine surgery in our institution. The CT scans were performed with the Mindways QCT calibration phantom placed beneath the patient. The CT volume data were then transferred to the QCT workstation to measure the BMD of vertebral body of L2, L3 and L4. Paired t test was performed to determine the difference between two groups with SPSS 15.0. Results It was no statistically significant in differences of BMD measured in 250 mA and 50 mA in lumbar spine L2-IA (P 〉 0.05). Radiation dose was 229.06±12.35 mGy.cm for 250 mA and 45.78±2.48 mGy.cm for 50 mA. Conclusion Low-dose CT scan with 50mA for lumbar spine QCT BMD measurement ensures the accuracy of BMD measurement, and decreases the radiation dose greatly.

关 键 词:骨密度 X线计算机 体层摄影术 定量CT 剂量 

分 类 号:R815[医药卫生—放射医学] R68[医药卫生—临床医学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象