全自动尿沉渣分析仪UF-1000i性能评价  被引量:9

Evaluation of performance of flow cytometer urine sediment analyser UF-1000i

在线阅读下载全文

作  者:张崇唯[1] 朱文彬[1] 金咏梅[1] 熊明[1] 谢恒[1] 邓山鹰[1] 粟军[1] 

机构地区:[1]四川大学华西医院实验医学科,四川成都610041

出  处:《现代预防医学》2013年第1期127-129,133,共4页Modern Preventive Medicine

摘  要:目的对全自动尿沉渣分析仪UF-1000i进行性能评价,判断其分析性能是否满足临床要求。方法按照国家实验室认可及CAP认可要求,检测UF-1000i检测指标数(RBC、WBC、EC、CAST、BACT)的精密度、准确度、携带污染率、线性范围并且验证参考区间。结果 UF-1000i测定各指标的日内、日间精密度和准确度均在仪器要求范围内,并具有良好的线性和较低的携带污染率。结论实验室UF-1000i各项性能均符合仪器要求范围,也满足了CAP认可对性能评价要求,可应用于临床检测,并提供了持续性的质量保证。OBJECTIVE To evaluate the function of flow cytometer urine sediment analyzer UF-1000i in order to meet clinical requirements basically. METHODS The evaluation was performed according to guidelines established by accreditation of nation- al laboratory and CAP. Precision, accuracy, carryover, linearity were analyzed and the reference interval was verified. RE- SULTS The with-day and between-day precision, accuracy and reference interval of UF-1000i were in the allowable range and the results by UF-1000i also possessed good linearity and the carryover was lower. CONCLUSION All properties of UF-1000i in laboratory meet standard requirements, also satisfy the demand of CAP laboratory accreditation, we can extensively use them in clinic and provide a continuous quality assurance.

关 键 词:CAP实验室认证 尿沉渣分析仪 性能评价 精密度 准确度 