利用模拟定位机验证射波刀影像同步追踪系统的准确性研究  被引量:3

Accuracy of Simulator Verification Cyber Knife Image Guided and Synchrony Tracking System

在线阅读下载全文

作  者:曹洋森[1] 于春山[1] 刘永明[1] 

机构地区:[1]上海长海医院放疗科,上海200433

出  处:《科学技术与工程》2013年第6期1578-1580,共3页Science Technology and Engineering

摘  要:利用模拟定位机验证射波刀影像引导系统及同步追踪系统准确性的可行性。用Synchrony模体及Xsight Lung模分别至于射波刀及模拟定位机下,测量各自环境下模体的运动数据并进行比较。结果Synchrony模体与Xsight Lung模体在拟机下测得的数据与射波刀影像追踪系统分析得出的运动数据误差分别为0.1 mm和0.9 mm。说明利用模拟机测量模体动数据与射波刀影像追踪系统分析所得数据相符,但需要建立完善的呼吸模型。The accuracy of the image-guided system and synchronous tracking system of CyberKnife was verfied. Synchrony Phantom and Xsight Lung Phantom were placed the CyberKnife and Simulator, then the data of motion in each environment were measured and compared. It is resulted the motion data errors were 0. 1 mm and 0. 9 mm of Synchrony and Xsight Lung Phantom between CyberKnife image tracking system and Simulator. It is conclused the data of simulator motif motion is consistent with CyberKnife image tracking system, but need to establish a comprehensive respiratory model.

关 键 词:射波刀 模拟定位机 影像引导 同步追踪 

分 类 号:R811.1[医药卫生—放射医学]

 