UF-50尿沉渣分析仪测定尿液红细胞的干扰因素  被引量:1

UF-50 Urinary Sediment Analyzer Test Urine Erythrocytic Interference Factors

在线阅读下载全文

作  者:董青兰 张国伟 张伟 

机构地区:[1]山西省侯马市人民医院检验科,043000

出  处:《临床合理用药杂志》2013年第6期99-99,共1页Chinese Journal of Clinical Rational Drug Use

摘  要:目的探讨UF-50全自动尿沉渣分析仪测定尿液中红细胞的影响因素。方法随机样本分别用UF-50和显微镜检测尿中红细胞。结果 UF-50测定红细胞阳性率(12.5%)高于镜检(9.1%),二者符合率为72%。结论结晶、霉菌、滴虫、脂肪滴、精子及大量细菌导致UF-50检测红细胞时出现假阳性,应显微镜复检。Objective To study the UF -50 automatic urinary sediment analyzer measurement in the urine, the influ- ence factors of red blood cells. Methods Random sample respectively with UF - 50 and microscope detection urine - determina- tion of red blood cells. Results The UF -50 measurement erythrocyte positive rate ( 12. 5% ) than microscopy (9. 1% ), the coincidence rate is 72%. Conclusion Crystal, mould, trichomonad, fat granule, sperm and a large number of bacteria that cause UF- 50 red blood cell detection occurs when the false positive, should microscope review.

关 键 词:红细胞 显微镜检查 UF-50尿沉渣分析仪 

分 类 号:R446.12[医药卫生—诊断学]

 