检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
机构地区:[1]安徽医科大学附属省立医院骨二科,安徽省合肥市230001
出 处:《中国组织工程研究》2013年第4期594-598,共5页Chinese Journal of Tissue Engineering Research
摘 要:背景:在全髋关节置换中,髋臼缺损很常见,为重建一个稳定、牢固的髋臼,置换过程中往往需要髋臼的重建。目的:探讨螺钉固定钛网结合骨水泥型髋臼杯在髋臼重建中应用的早期疗效。方法:纳入采用螺钉固定钛网结合骨水泥型臼杯重建髋臼髋关节翻修的患者23例。疼痛病史1-3年,术前Harris评分20-48分,平均34分。重建后采用X射线平片评价髋臼假体松动标准进行影像学评价及Harris评分定期随访。结果与结论:重建后随访1-3年,平均随访24个月,无脱失。重建后3个月Harris评分较重建前显著增高,置换后2年髋关节活动度较置换前显著增高,差异均有显著性意义(P<0.05)。2例患者下肢不等长,1例患者下蹲后出现脱位,保守治疗后好转。最后一次随访时无早期感染、脱位,无盆腔不适感。重建后随访影像学上显示无髋臼假体松动与移位,未发生钛网断裂。说明螺钉固定钛网结合骨水泥型髋臼杯治疗65岁以上髋关节翻修患者其近期疗效十分满意,后期疗效还有待于进一步随访观察。BACKGROUND: In total hip arthroplasty, acetabular defects are common, for the reconstruction of a stable and solid acetabulum, the acetabular reconstruction is often required during the replacement process.OBJECTIVE: To investigate the early effect of the short term results of titanium mesh fixed by screws combined with a cemented acetabular cup in acetabular reconstruction.METHODS: Twenty-three patients treated with acetabular reconstruction by screw fixed titanium mesh combined with a cemented acetabular cup wereselected. The pain history was 1 to 3 years. Harris scores before operation were 20-48 with an average score of 34. X-ray plain film was used to evaluate theacetabular prosthesis loosening standards after reconstruction for the radiographic evaluation and Harris score of regular follow-up.RESULTS AND CONCLUSION: All the patients were followed-up for 1-3 years and average in 24 months without losing. The Harris score at 3 months after operation was higher than that before operation, and the hip activity at 2 years after operation was higher than that before operation, and the difference was significant (P 〈 0.05). Two patients still had leg length discrepancy, one patient with dislocation after squatting, and recovered after conservative treatment. The last follow-up showed there were no early infection, dislocation and pelvic discomfort. Radiographic results showed that there was no aseptic loosening and displacement of the acetabular cap and there was no fracture on titanium mesh. Screw fixed titanium mesh combined with a cemented cup is satisfactory process in the treatment of hip revision in over 65 years old patients, and the long-term effects are needed to be further followed-up and observed.
关 键 词:骨关节植入物 人工假体 骨水泥型髋臼杯 钛网 髋关节 髋臼假体 髋臼缺损 髋臼重建 假体置换 螺钉 内固定 髋关节翻修
分 类 号:R318[医药卫生—生物医学工程]
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.28