作 者:张士胜[1,2] 张静 姚文琴 廖华萍 戴南平 于青 刘春雷 董世奇 廉井财[1,2]
机构地区:[1]上海瑞视眼科,上海市200011 [2]上海交通大学医学院附属瑞金医院眼科
出 处:《中国激光医学杂志》2013年第2期78-81,共4页Chinese Journal of Laser Medicine & Surgery
基 金:上海市重点学科建设项目资助(S30205)
摘 要:目的比较Pentacam系统与A型超声角膜测厚仪测量准分子激光手术前近视患者中央角膜最薄点厚度的差异。方法对准备行准分子激光角膜屈光手术的近视患者69例138只眼,分别采用Pentacam系统和A型超声角膜测厚仪测量角膜中央最薄点厚度,对测量结果进行统计学分析。结果 Pentacam系统、A型超声角膜测厚仪测量中央角膜最薄点厚度值分别为(548.0±28.9)μm和(538.8±29.4)μm,两者高度正相关(P<0.05)。Pentacam系统比A型超声测得的中央角膜厚度厚(9.2±9.9)μm,差异有显著意义(P<0.05),该差值与A型超声测量角膜厚度值负相关(P<0.05),而与屈光度、年龄等无关。结论 Pentacam测量近视患者角膜厚度值略高于A型超声法,尽管二者有较好的一致性,但相互间还不能完全替代。Objective To compare the difference in the thinnest central corneal thickness (CCT) measurements in myopic eyes obtained with Pen- tacam and A-scan ultrasound pachymetry respectively. Methods The thinnest CCT was determined in 138 eyes of 69 myopic patients before excimer-laser refractive surgeries first with Penta- cam, then an A-scan ultrasonic pachymetry. The differences in the measurements obtained via two methods were statistically analyzed.Resular The thinnest CCT readings with Pentacam and ultrasonic pachymetry were (548.0 ±28. 9) μm and (538. 8±29. 4) p,m re- spectively. The correlation between them was high (P 〈 0. 05 ). There was a significant difference between the values obtained with the two methods (9. 2±9.9) μm, (P 〈 0. 05 ). The difference negatively correlated with the CCT obtained with A-scan ultrasonic pachym- etry(P 〈0. 05), while having nothing todo with such factors as refractive power and age. Conchlusions For myopic eyes, pentacam tends to give a bit higher CCT reading than ultrasound pachymetry. Despite of a relatively good agreement between them, they cannot be replaced completely by each other.
关 键 词:激光 PENTACAM系统 超声测厚仪 角膜厚度 近视
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:18.190.158.12