流动注射-化学发光法测定血浆中盐酸阿霉素  被引量:4

Flow injection chemiluminescence determination of doxorubicin hydrochloride in plasma

在线阅读下载全文

作  者:葛闯[1] 付虎[1] 何云[1] 易钢[1] 

机构地区:[1]重庆医科大学临床检验诊断学教育部重点实验室,重庆400016

出  处:《分析试验室》2013年第6期105-107,共3页Chinese Journal of Analysis Laboratory

摘  要:在碱性介质中,K3[Fe(CN)6]与鲁米诺产生化学发光反应,盐酸阿霉素对该体系的发光强度具有显著的抑制作用,且光信号的减小值(△I)与盐酸阿霉素的浓度在一定的范围内成正比。基于此建立了一种流动注射化学发光法测定盐酸阿霉素的新方法。在最优实验条件下,该方法的线性范围为0.4~2.5μg/mL,检出限为2.9×10-2μg/mL。对1.2μg/mL的盐酸阿霉素连续平行测定11次,其相对标准偏差为2.1%。该方法用于血浆中盐酸阿霉素的分析,回收率在98.4%~102.0%之间。A new analytical method coupled with the flow injection technique for the detection of doxorubicin hydrochloride(DOX) was developed based on the fact that the chemiluminescence reaction of luminol and potassium ferricyanide could be suppressed by DOX in alkaline media, and the inhibitory intensity was linear with the concentration of DOX. The parameters which affect the chemiluminescence signal were investigated. Under the optimized conditions ,the linear range of 0. 4-2. 5μg/mL with the detection limit of 2.9 -10^-2 μg/mL was gained for the DOX. The relative standard deviation (RSD) of 2. 1% was obtained for eleven-time consecutive determinations of 1.2μg/mL DOX. The method was applied to the detection of DOX in plasma, and the recoveries were in the range of 98.4% - 102. 0%.

关 键 词:抑制化学发光 流动注射 鲁米诺 铁氰化钾 血浆 盐酸阿霉素 

分 类 号:O657.3[理学—分析化学]

 