检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
作 者:刘芳[1] 邹建中[1] 艾慧坚[1] 焦娇[1] 赵欣[1]
机构地区:[1]重庆医科大学生物医学工程学院省部共建超声医学工程国家重点实验室超声医学工程重庆市市级重点实验室,重庆400016
出 处:《中国介入影像与治疗学》2013年第6期369-372,共4页Chinese Journal of Interventional Imaging and Therapy
基 金:国家重点基础研究发展计划(973计划)项目(2011CB707902);国家自然科学基金重点项目(30830040);国家自然科学基金面上项目(30970827)
摘 要:目的以离体猪肝脏灌流模型探讨脉冲高强度聚焦超声(pHIFU)辐照下血流对热沉积的影响。方法采用pHIFU以定点方式辐照离体再灌注猪肝组织和非再灌注猪肝,辐照功率为80 W,辐照深度20mm,脉冲重复频率为100Hz。占空比为40%、30%、20%,辐照时间均为20s。于每组参数下重复辐照10次,共60个辐照点(再灌注组30个辐照点与非再灌注组30个辐照点)。辐照中使用热电偶测量靶区温度。辐照后即刻采集B超声像图,计算灰度增强区域的面积。辐照结束后,将猪肝切成厚为1~2mm薄片,找出最大损伤所在层面,测量猪肝实际坏死体积。结果各参数辐照下,离体再灌注猪肝中靶区温升均慢于非再灌注猪肝,且平均最高温度低于非灌注猪肝(P<0.05)。辐照后即刻,pHIFU辐照再灌注所致声像图中灰度增强区域的面积小于非再灌注猪肝(P<0.05),但平均灰度变化值差异无统计学意义。辐照结束后,pHIFU所致再灌注猪肝中的坏死体积小于非灌注猪肝。结论灌流条件下,血流循环的存在使pHIFU辐照所致血管周围猪肝组织中的热沉积减小。Objective To observe the impact on thermal deposition for in vitro pork liver exposed to pulsed high intensity focused ultrasound (pHIFU). Methods A circulated and a nowcirculated pork liver were exposed to pHIFU for 20 s at a 20 mm depth, respectively, pHIFU was maintained at 80 W acoustic power, 100 Hz pulse repetition frequency. The duty cycles were 40%, 30% and 20%. After 10 repeated pHIFU irradiations under every parameter, all 60 exposures were per-formed, with 30 in the circulated liver and the other 30 in the non-circulated liver. During pHIFU exposures, the tempera tures in the focal region were detected using a thermocouple embedded into the liver. B-mode ultrasonic images were also collected at the moment after irradiations, and the area of hyperecho regions were calculated. After all irradiations, the liv-ers were sliced into 1-2 mm, and the maximum lesions were photographed. Then, the volumes of the pictured necrosis were computed. Results During all pHIFU exposures, compared to non-circulated liver, the temperature raised slowly and the mean peak temperatures were significantly smaller in the circulated liver (P〈0.05). After the circulated liver exposed to pHIFU, less hyperecho regions and necrosis area were observed in the circulated liver (P〈0.05). However, the values of increased echoes in B-mode ultrasonic images were very nearly the same after the two liver irradiations. Conclusion The thermal deposition in pork liver tissues around vessels decreases because of circulation of blood flow during pHIFU irradia- tion.
关 键 词:高强度聚焦超声消融术 灌流 肝脏 凝固性坏死
分 类 号:R318.51[医药卫生—生物医学工程]
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.117