迭代重建算法前瞻性心电门控大螺距肺动脉双源CT成像的图像质量  

Image quality of prospectively ECG-triggered high-pitch dual source CT using iterative reconstruction in pulmonary angiography

在线阅读下载全文

作  者:孙喜霞[1] 季颖群[2] 郑飞[1] 赵承琳[1] 贾崇富[1] 杨志强[1] 卢晓娟[1] 刘兴众[1] 王照谦[1] 

机构地区:[1]大连医科大学附属第一医院心脏CT检查科,辽宁大连116011 [2]大连医科大学附属第一医院呼吸科,辽宁大连116011

出  处:《中国介入影像与治疗学》2013年第6期376-380,共5页Chinese Journal of Interventional Imaging and Therapy

基  金:国家"十二五"科技支撑计划课题项目(2011BAI11B15)

摘  要:目的与滤波反投影(FBP)算法比较,探讨基于原始数据的迭代重建(SAFIRE)算法大螺距肺动脉双源CT(DSCT)成像的图像质量。方法对临床怀疑肺栓塞的69例患者行前瞻性心电门控大螺距肺动脉CT扫描,分别采用SAFIRE和FBP算法进行图像重建,比较两种重建算法的肺动脉CT图像质量、噪声、SNR和CNR。结果 SAFIRE算法图像的肺动脉CT图像质量显著优于FBP算法[图像质量评分分别为(1.28±0.36)分和(1.99±0.56)分,P<0.001];SAFIRE算法的噪声显著低于FBP算法[分别为(14.60±2.11)HU和(18.52±2.65)HU,P<0.001];SAFIRE算法的SNR和CNR显著高于FBP算法(SNR分别为26.57±4.99和20.23±3.94,P<0.001;CNR分别为23.41±4.58和17.77±3.62,P<0.001)。结论前瞻性心电门控大螺距肺动脉DSCT成像时,与FBP算法相比,SAFIRE算法的肺动脉CT图像噪声更低,图像质量更高,显示肺动脉细小分支更佳。Objective To evaluate the image quality of prospectively ECG-triggered high-pitch dual-source CT (DSCT) in pulmonary angiography using sinogram affirmed iterative reconstruction (SAFIRE) compared with filtered back projection (FBP). Methods Totally 69 patients with suspected pulmonary embolism underwent prospectively ECG-triggered high-pitch DSCT in pulmonary angiography. The images were reconstructed with SAFIRE and FBP. Image quality, image noise, SNR and CNR were compared between the two methods. Results Image quality of SAHRE was significantly better than FBP (image quality scores was 1.28±0.36 and 1.99±0.56, P〈0. 001). The image noise of SAFIRE was signifi-cantly lower than FBP (image noise was [14.60 ± 2.11] HU and [-18.52 ± 2.65] HU, P〈0. 001). SNR and CNR of SAFIRE were significantly higher than FBP (SNR was 26.57±4.99 vs 20.23±3.94, P〈0. 001; CNR was 23.41±4.58 vs 17.77±3.62, P〈0. 001). Conclusion Compared with FBP, prospectively ECG-triggered high-pitch DSCT pulmonary angiography using SAFIRE can significantly reduce image noise and improve image quality, especially for displaying subdi- vision arteries.

关 键 词:肺动脉 体层摄影术 X线计算机 图像质量 迭代重建 

分 类 号:R816.4[医药卫生—放射医学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象