Sysmex SE-9000血液分析仪白细胞分类结果的评价  被引量:1

Evaluation on Classification of Leukocytes by Sysmex SE-9000 Hematology Analyzer

在线阅读下载全文

作  者:黄劲[1] 张敬[1] 彭政 

机构地区:[1]同济医科大学附属同济医院检验科,武汉430030

出  处:《同济医科大学学报》2000年第4期374-375,共2页Acta Universitatis Medicinae Tongji

摘  要:将 Sysm ex SE- 90 0 0血液分析仪 5项分类结果与手工分类结果进行了分析与比较。根据血液分析仪的各种提示分为阴性组与阳性组。运用 Excel对分类结果进行统计和分析 ,结果发现 :无论是阴性组还是阳性组 ,中性粒细胞、淋巴细胞、嗜酸粒细胞两种方法之间相关性非常显著 (r=0 .912 ) ,而嗜碱性粒细胞、单核细胞有极显著性差异 (P<0 .0 1) ,且相关性差 (r=0 .5 2 3)。提示当单核细胞 ,嗜碱粒细胞高于正常范围时 ,应镜检为宜。The differential leukocyte counts by Sysmex SE 9000 Hematology Analyzer were compared with those of blood smears under microscopy. The samples were divided into negative group and positive group according to the warning of the analyzer. The classifying results were analyzed by means of software Excel. The results showed that the relatives of two methods for neutrophils, lymphocytes and eosinophils were remarkable ( r =0.912). There was significant difference in basophils and monocytes between the two groups ( P <0.01) with weak relatives ( r =0.523). It was suggested that when the count of monocytes and basophils is out of range, it is better to analyze under microscopy.

关 键 词:SysmexSE-9000血液分析仪 白细胞分类 评价 

分 类 号:R446.11[医药卫生—诊断学]

 