新型芯片传感-微分电位溶出法测定血铅的探讨  被引量:1

Study on the new sensor chip-differential potentiometric stripping analysis in the detection of blood lead

在线阅读下载全文

作  者:唐寅 盛青松 张同军 辛君伟 

机构地区:[1]张家港市中医医院检验科,江苏张家港215600 [2]无锡市申瑞生物制品有限公司,江苏无锡214092

出  处:《检验医学》2013年第8期704-706,共3页Laboratory Medicine

摘  要:目的评价芯片传感-微分电位溶出法在临床血铅检测的准确度和重复性,并与氢化物发生原子荧光法比较,为临床选择检测方法提供参考依据。方法分别运用芯片传感-微分电位溶出法和氢化物发生原子荧光法同时检测血液样本和标准物质(为2011年全国血铅检测室间质评样本),评价芯片传感-微分电位溶出法的正确度、重复性、回收率、可报告范围和符合率。结果以靶值±10%作为允许误差,6种标准物质的芯片传感-微分电位溶出法测定结果与靶值比较,均在允许误差范围之内,正确度良好;低、中、高3个水平样本的相对标准偏差(RSD)分别为6.98%、4.09%、2.78%;平均回收率为98.1%,系统比例误差为1.9%;与氢化物发生原子荧光法比较,差异无统计学意义(P>0.05),线性回归分析相关系数(r2)=0.964,2种方法测定结果的符合度为96.4%。结论芯片传感-微分电位溶出法准确度高、重复性好、精密度佳,适用于临床血铅检测。Objective To evaluate the accuracy and repeatability of clinical blood lead detection by sensor chip- differential potentiometric stripping analysis and compare with hydride generation atomic fluorescence spectrometry, and to provide the reference for clinical application. Methods The 2 methods were used to analyze the blood samples and the certified reference materials ( national blood lead detection external quality assessment samples in 2011 ). The accuracy, repeatability, recovery rate, acceptable range and coincidence rate of sensor chip-differential potentiometric stripping analysis were evaluated. Results All of the results of 6 certified reference materials were in the allowable error range(target value + 10% ), and the accuracy was good. The relative standard deviations (RSD) of the low, medium and high blood lead level blood samples were 6.98% ,4.09% and 2.78% , respectively. The average recovery rate was 98.1% , and the system ratio error was 1.9%. Compared the results of sensor chip-differential potentiometric stripping analysis and hydride generation atomic fluorescence spectrometry, the results had no statistical significance ( P 〉 0.05 ) , and the correlation coefficient of linear regression analysis ( rE ) was O. 964. The coincidence rate of the 2 methods was 96.4%. Conclusions sensor chip-differential potentiometric stripping analysis has good accuracy, repeatability and precision, and is suitable for the clinical application of detecting blood lead.

关 键 词:血铅 芯片传感器 微分电位溶出法 氢化物发生原子荧光法 

分 类 号:R446.1[医药卫生—诊断学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象