出 处:《Optoelectronics Letters》2013年第5期385-388,共4页光电子快报(英文版)
基 金:supported by the National Natural Science Foundation of China(No.61077074);the National Key Technology R&D Program(No.2013BAK06B04);the Science and Technology Department of Jilin Province of China(No.20120707);the Science and Technology Development of Changchun City of China(No.11GH01)
摘 要:Based on wide-band infrared(IR) light source and dual-channel pyroelectric detector with detection channel of 4.66 μm and reference channel of 3.95 μm,a differential mid-infrared(MIR) carbon monoxide(CO) detector is designed and implemented.In order to reduce the detection limit and improve the detection sensitivity,an open spherical mirror chamber is designed and fabricated according to the divergence angle of the light source.The CO detection system is established using the welded and debugged detection circuits,a series of CO gases with different concentrations are prepared,and gas concentration calibration experiment is carried out.Experimental results indicate that after the amplifying circuit,the signal-to-noise ratios(SNRs) of the two channels are 17.58 dB and 18.46 dB,respectively,and the detection error of this system is less than 9% in 0%-4% measuring range.The detection sensitivity in the low concentration range is approximately 0.05%.6 h measurement on the 0 ppm sample gas shows that the fluctuation range is about ±0.02%,and the measurement standard deviation is about 0.89%.Based on wide-band infrared (IR) light source and dual-channel pyroelectric detector with detection channel of 4.66 μm and reference channel of 3.95μm, a differential mid-infrared (MIR) carbon monoxide (CO) detector is designed and implemented. In order to reduce the detection limit and improve the detection sensitivity, an open spherical mirror chamber is designed and fabricated according to the divergence angle of the light source. The CO detection system is established using the welded and debugged detection circuits, a series of CO gases with different concentrations are prepared, and gas concentration calibration experiment is carried out. Experimental results indicate that after the amplifying circuit, the signal-to-noise ratios (SNRs) of the two channels are 17.58 dB and 18.46 dB, respectively, and the detection error of this system is less than 9% in 0%-4% measuring range. The detection sensitivity in the low concen- tration range is approximately 0.05%. 6 h measurement on the 0 ppm sample gas shows that the fluctuation range is about±0.02%, and the measurement standard deviation is about 0.89%.
关 键 词:一氧化碳 中红外 检测仪 设计 检测灵敏度 CO气体 检测系统 气体浓度
分 类 号:TN214[电子电信—物理电子学]
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...