5种光合有效辐射表性能测试  被引量:1

Performance Test of Five-type Photosynthetic Active Radiometers

在线阅读下载全文

作  者:丁蕾[1] 杨云[1] 边泽强[1] 权继梅[1] 王冬[1] 

机构地区:[1]中国气象局气象探测中心,北京100081

出  处:《应用气象学报》2013年第5期617-624,共8页Journal of Applied Meteorological Science

摘  要:为认识和提高光合有效辐射(波长为400~700nm)的测量精度,对国内外5种常用的光合有效辐射表的灵敏度、余弦响应、非线性、温度依赖性、光谱响应等性能进行了测试和分析。结果表明:5种光合有效辐射表(简称光合表)的非线性误差均在1%以内;3种光合表(PARLITE,LI-190和FS-PR)的余弦误差均小于10%,温度系数均小于0.3%/℃,2种光合表(TRT一5和HSC—FPH-1)的余弦误差明显偏大,温度系数小于0.5%/℃。2种光合表(PARLITE和LI-190)的光谱响应曲线接近于理想光谱,而其余3种光合表与理想光谱曲线偏离较大。在自然光下的一致性比对测量结果显示,5种光合表测量的辐射量与标准值的误差均在5%以内,但综合评价后认为,3种光合表(PARLITE,LI-190和FS-PR)的性能明显优于其他2种光合表(TRT-5和HSC—FPH-1)。该结果可为气象观测业务选型提供参考。Photosynthetic active radiometers are used to measure the solar photosynthetic active radiation. Also, it can be turned 180°, facing down to measure the photosynthetic active radiation of the ground. When it is used outdoors, the environmental condition has impacts on the performance, which may increase the meas- urement error. The performances of five different types of photosynthetic active radiometers are tested, such as sensitivity, cosine response, linearity, temperature dependence, spectral sensitivity and non-stabil- ity. The non-linearity errors of five photosynthetic active radiometers are all within 1°. When the irradi- ance is less than 250 W · m^-2 , the differences of the non-linearity errors for different-type radiometers are significant. Cosine response errors gets bigger with the increase of the zenith angle (%). When the zenith angle changes from -80° to +80°, cosine response errors of PAR LITE, LI-190 and FS-PR type radiome- ters are within 10%; the error of TRT-5 radiometer is within 10% just in the range of -55° to °55°; and HSC-FPH-1 is within 10% when 0 is --45° to +40°. When testing the effects of temperature on photosyn thetic active radiometer, the temperature coefficients of PAR LITE, LI-190 and FS-PR type radiometers are within 0.3%/℃, TRT-5 and HSC-FPH-1 type photosynthetic active radiometers are within 0.5 %/℃. The wavelength range of photosynthetic active radiometer is from 400 nm to 700 nm. The ideal sensor should have clear cutoffs to light below 400 nm and above 700 nm. The spectral response of PAR LITE, LI-190 type radiometers is approximated to a horizontal line, which is closer to the ideal spectral curve. The response of FS-PR is overestimated within 400-550 nm and underestimated after that. The response of TRT-5 and HSC-FPH-1 photosynthetic active radiometers is lower than the ideal curve in the whole vis- ible waveband. The consistency of five-type photosynthetic active radiometers with the standard value are all within 5 % which is tested under solar

关 键 词:光合有效辐射表 余弦响应 光谱响应 性能测试 

分 类 号:P412.14[天文地球—大气科学及气象学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象