ICP-MS测定溴、碘、砷和硒在多目标样品中的应用  被引量:7

Determination of Bromine,Iodine,Arsenic and Selenium in Multi-purpose by ICP-MS

在线阅读下载全文

作  者:安国荣[1] 张启云[1] 陈辉[1] 魏振宏[1] 杜作朋 张莹[1] 

机构地区:[1]青海省地质矿产测试应用中心,西宁市盐湖巷15号810008

出  处:《光谱实验室》2013年第6期3306-3308,共3页Chinese Journal of Spectroscopy Laboratory

摘  要:样品经熔剂半熔后,水提取,用强酸性阳离子树脂除去大部分阳离子,在弱碱性介质中用ICP-M S测定溴、碘、砷、硒,样品分析结果与多目标推荐分析方法比对,相对偏差小于6.9%,另选取国家土壤一级标准物质12件对本方法的检出限、精密度、准确度进行测试,各项质量指标均能满足规范要求。Bromine, iodine, arsenic and selenium was determined by ICP-MS with sample pretreatment using semi-molten flux, water extraction and removing most cation using strong-acid cation exchange resin. Relative deviation is less than 6.9% of sample analysis by comparing muhi-objective recommended analysis method. The detection limit, precision, accuracy test of the method meet the requirements of the code by determined 12 national standard reference materials.

关 键 词:    电感耦合等离子体质谱仪 青海多目标地球化学调查 

分 类 号:O657.63[理学—分析化学]

 