注射泵流量误差校准规范的探讨  被引量:9

Discussion on Calibration Specifications for Syringe Pump Flow Rate

在线阅读下载全文

作  者:倪少石 倪萍[1] 马继民[1] 张鲁闽[1] 

机构地区:[1]南京军区福州总医院医学工程科,福州350025

出  处:《医疗卫生装备》2013年第10期107-108,115,共3页Chinese Medical Equipment Journal

摘  要:目的:通过对注射泵流量误差校准规范的分析和讨论,探讨产生流量误差的原理以及不同规范的检测侧重点。方法:对医用注射泵在重新设计的流速流量校准点所得到的检测数据进行分析。结果:低速流量点可以完全反映注射泵的准确度。流速设置值越高,实际流速越稳定;反之,流速波动越大。结论:目前,军队规范中仅测量低速流量测试点(5 mL/h)的方法是可行的。Objective To analyse and discuss the method in flow rate of calibration specifications for the safety and effectiveness of medical electrical equipment.Methods The test data measured at different flow rate points were analyzed.Results The accuracy of syringe pump could be reflected by the percent error measured at the lowest rate.Conclusion The higher the flow rate is,the smaller the percent error can be,and vice versa for the lower flow rate.The current calibration specification for army hospital,which only tests low flow rate point(5 mL/h),is feasible.

关 键 词:注射泵 质量控制 流速流量 准确度 

分 类 号:TH789[机械工程—仪器科学与技术]

 