模拟体外循环下肝素涂层的稳定性实验  

Study of heparin stability on heparin-coated circuits under simulating cardiopulmonary bypass

在线阅读下载全文

作  者:杨春[1] 杨剑[1] 罗鹏[1] 曹瑞军[2] 周华[2] 

机构地区:[1]第四军医大学西京医院心血管外科,陕西西安710032 [2]西安交通大学应用化学系,陕西西安710049

出  处:《心脏杂志》2013年第5期573-575,共3页Chinese Heart Journal

基  金:国家自然科学基金项目资助(30700174);陕西省社发攻关基金[(2006K12-G5(4)]项目资助

摘  要:目的:评估肝素涂层体外循环管道肝素的稳定性。方法:测定肝素图层体外循环管道模拟体外循环法与浸泡法在不同时点的肝素脱落率。结果:在128 h内,两组肝素结合表面均有不同程度的肝素分子脱落现象,肝素脱落率随时间增大,在一定时间内浸泡法结合的肝素表面的肝素最大脱落率为85%,而体外循环法结合表面的肝素脱落率为92%。体外循环组涂层表面中肝素的脱落率与浸泡组两者之间无显著差异。结论:短期内肝素结合体外循环管道比较牢固,满足临床使用。AIM: To evaluate heparin stability on heparin-coated cardiopulmonary bypass (CPB) circuits. METHODS: Heparin off-rates of CPB stimulation method and immersion method at different time points were determined. RESULTS: At 128 h, different levels of heparin molecule shedding phenomenon were observed in heparin-binding surface in both groups and the heparin off-rate increased with time. The maximum heparin off-rates of the CPB method and immersion method were, respectively, 92 and 85 during a certain period of time, with no significant difference between groups. CONCLUSION: Heparin stability on heparin-coated circuits of CPB has relatively stronger heparin binding and is of better clinical use.

关 键 词:肝素涂层管道 脱落率测定 肝素稳定性 

分 类 号:R608[医药卫生—外科学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象