检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
机构地区:[1]武汉大学中南医院放化疗科,湖北武汉430071 [2]武汉市普爱医院肿瘤内科,湖北武汉430033
出 处:《中国医疗设备》2013年第12期124-126,共3页China Medical Devices
摘 要:目的探讨多叶准直器射野和独立准直器射野剂量学数据对放射治疗计划系统建模的影响。方法在Varian Unique型直线加速器上对多叶准直器和独立准直器射野的百分深度剂量、射野离轴量以及总散射因子进行比较。结果两种射野的百分深度剂量差别不大,但射野离轴量以及射野总散射因子差别较大。多叶准直器形成的射野均比独立准直器形成的射野要大;随射野变小,两者总散射因子的差异越来越显著。结论在计划系统的建模过程中必须关注不同射野的剂量学数据差异,建议在调强计划系统建模过程中采用多叶准直器射野进行数据采集。Objective To evaluate the influence of dosimetry data of radiation fields formed by multileaf collimator and independent collimator on the commissioning of treatment planning system. Methods The percentage depth dose, the off-axis ratio and the total scattering factors of radiation fields formed by multileaf collimator and independent collimator were measured in Varian Unique linac respectively. Results There is no significant difference in the percentage depth dose of two kinds of fields. There are significant differences in the off-axis ratio and the total scattering factors of two kinds of fields. The sizes of radiation fields formed by multileaf collimator are larger than the sizes of radiation fields formed by independent collimator. The differences in the total scattering factors of two kinds of fields are more and more significant along with the decrease of field sizes. Conclusion We must pay close attention to the differences of dosirnetry data of radiation fields formed by multileaf collimator and independent collimator in the commissioning process of treatment planning system, which indicates that we should collect dosimetry data by using fields formed by multileaf collimator in the commissioning process of intensity-modulated treatment planning system.
关 键 词:多叶准直器 独立准直器 照射野 放射治疗计划系统
分 类 号:TL53[核科学技术—核技术及应用] TH774[机械工程—仪器科学与技术]
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.15