RQR5 X射线辐射质的产生及空气比释动能不确定度的评定  被引量:2

Medical Diagnostic X-ray Equipment Radiation Condition RQR5 and Its Uncertainty Evaluation of Air Kerma

在线阅读下载全文

作  者:陆建荣[1,2] 刘有军[1] 于苒[1] 刘荣黎[1] 

机构地区:[1]北京工业大学生命科学与生物工程学院,北京100124 [2]中国人民解放军总后勤部卫生部药品仪器检验所,北京100071

出  处:《北京工业大学学报》2013年第12期1916-1920,共5页Journal of Beijing University of Technology

基  金:国家自然科学基金资助项目(11172016)

摘  要:针对目前诊断水平剂量计缺少校准装置的现状,采用X射线机、定位装置和温度气压测量仪等组成标准辐射场,建立诊断水平剂量计校准装置;通过对辐射场的空气比释动能进行测量和分析,得出其相对不确定度U=3.8%(k=2);此标准辐射场能满足诊断水平剂量计的校准.In view of the current situation of lacking diagnostic level dosimeter calibrating unit, a standard radiation condition composed of X-ray equipment, positioning device, temperature measuring instrument and barometric pressure measuring instrument was provided, thus building the standard condition for the calibration of diagnostic level dosimeter. The uncertainty of this radiation condition was evaluated as U = 3.8% (k = 2) by measuring and analyzing the air kerma of the radiation field. This standard radiation condition can meet the calibration of diagnostic level dosimeter.

关 键 词:RQR5 空气比释动能 不确定度 