基于多次反射样品池的TDLAS逃逸NH_3检测研究  被引量:6

Detection of escaping NH_3 based on TDLAS with multi-pass cells

在线阅读下载全文

作  者:张增福[1] 邹得宝[1] 陈文亮[1] 赵会娟[1] 徐可欣[1] 

机构地区:[1]天津大学精密测试技术及仪器国家重点实验室,天津300072

出  处:《光电子.激光》2013年第12期2367-2371,共5页Journal of Optoelectronics·Laser

基  金:国家"863"计划(2012AA022602);国家重大科学仪器专项(2012YQ 060165)资助项目

摘  要:时域激光吸收光谱(TDLAS)技术具有高选择性、高准确性优点,为逃逸NH3的在线检测提供了可靠的技术手段。本文利用TDLAS系统分别在White和Herriott两种样品池中对浓度为(10~100)×10^-6的NH3进行检测,采集得到其二次谐波光谱,比较了两种样品池测量结果,分析对比两种样品池的优缺点。实验显示,Herriott样品池的噪声要大,根据光谱曲线的标准偏差得到White样品池最低检出限为1.21×10^-6,Herriott样品池的检测限为1.61×10^-6。结果表明,Herriott样品池更适合恶劣环境现场应用,能够满足逃逸NH3的痕量检测要求,适合在线监测。Time-domain laser absorption spectroscopy (TDLAS) with the advantages of high selectivity and high accuracy provides a reliable means of on-line detection technology for escaping ammonia (NH3). The concentrations from 10 ×10^-6 to 100× 10^-6 of NH3 were detected vdth the TDLAS system in the two multi-pass cells of White cell and Herriott cell, and their second harmonic spectra were collected. The two kinds of cell measurement results are compared, and the advantages and disadvantages are analyzed and compared. Experiment results show that the noise of the Herriott cell is higher than that of the White cell. According to the standard deviation of spectral curves, the detection limit of NH3 concentration is 1.21 × 10^-6 in White cell,and is 1.61 × 10^-6 in Herriot cell. The results show that the Herriott cell is more suitable for the harsh environment, and the system can meet the requirements of trace detection of escaping ammonia, suitable for on-line monitoring.

关 键 词:时域激光吸收光谱(TDLAS) NH3气检测 White样品池 Herriott样品池 

分 类 号:O443.1[理学—电磁学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象