机构地区:[1]上海交通大学医学精密工程及智能系统研究所,上海200240
出 处:《北京生物医学工程》2013年第6期624-631,636,共9页Beijing Biomedical Engineering
基 金:国家自然科学基金(30800235);上海交通大学医工(理)交叉研究基金(YC2010MS75)资助
摘 要:目的设计一种新型的植入式胃肠道刺激系统,不仅具有刺激功能,还具有肠电和压力检测功能,可用于检测胃肠道刺激的效果,同时增加无线能量供给,以实现刺激器的长期植入。方法系统由体内刺激模块、体外控制模块及无线能量传输模块组成。体外控制模块通过无线射频将控制信号传输到体内刺激模块,体内刺激模块的能量由体外能量发射装置通过电磁耦合进行供给。通过生物反馈控制检测不同刺激参数对胃肠道收缩活动的作用效果,实时调整刺激参数,输出需要的刺激脉冲。以模拟心电信号模拟肠电信号,进行了相关的体外实验。结果在体外实验中,系统可有效检测到2—20次/min的模拟心电信号,并实现实时刺激参数修改输出不同的刺激脉冲。该系统实现了电流检测功能,监测作用部分的胃肠电阻。经皮无线能量在两级线圈轴向距离为22mm时的接收充电稳定功率最大为0.93W,体内锂离子的充电电流为180~240mA。结论系统可检测到最大变化范围的模拟肠电信号。验证电流的作用效应为后续的恒流刺激模式提供参考。该系统的无线充电功能可满足植入式刺激器长期植入的能量需求。stimulating function Objective A new and the function of electrical stimulation. And the function type of implantable electrical stimulation system is designed with the intestinal signal and pressure detection in order to detect the result of of wireless energy supply is added to insure that the stimulator can be long-time planted. Methods The system consists of internal stimulator, external controller and wireless energy transmitting modules. The control signal is transformed from external controller to internal stimulator through RF. The energy of the internal stimulator is captured from external transmitting devices by using electromagnetic couple technology. We detect gastrointestinal contraction under different stimulating parameters with biofeedback control technology. So we can adjust the stimulating parameters to output stimulating pulse we need. Results This system detects 2 to 20 cpm analogous electrocardiograph signals effectively for adjusting the parameter to produce different stimulating pulse in vitro. Meanwhile, we realize current detecting function and the detection of the gastrointestinal resistor stimulated. The maximum of received stable power in vivo is 0.93 W and the charging current can be 180 mA to 240 mA as the coil distance is 22 ram. Conclusions This system can detect the maximum variation range of simulated intestinal signal. Verification of the current action effects provides areference for constant current stimulation mode in the future. The wireless charging function meets the energy requirement for the long-term implantation of the stimulator.
分 类 号:R318.6[医药卫生—生物医学工程]
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.28