作 者:范秋红[1] 邓振生[1] 欧阳丽蓉[2] 王小宜[2]
机构地区:[1]中南大学地球科学与信息物理学院生物医学工程研究所,湖南长沙410083 [2]中南大学湘雅医院放射科,湖南长沙410008
出 处:《中国医学影像技术》2014年第1期141-145,共5页Chinese Journal of Medical Imaging Technology
基 金:中南大学中央高校基本科研业务费专项基金(2012zzts070)
摘 要:目的研究不同测量方法及图像获取条件对胸腔气道测量的准确性。方法将亚克力圆管插入泡沫中制成气道模型,应用Siemens64排螺旋CT进行扫描,然后用不同重建内核对不同视野下获得的数据进行重建,得到气道模型CT图像;分别用半高宽法(FWHM)、二阶零交点方法(zC)和相位一致方法(PC)对气道模型CT图像进行测量,比较测量误差。结果对壁厚约为1.5~2.0mm的管道,当重建内核从B30f变化到B60f时,FWHM和ZC的测量误差从正值变为逐渐增大的负值,并在B45f时达到最小(〈11%),PC的测量误差相对稳定。对于壁厚约为1mm的管道,FWHM和ZC的测量误差在B30f时最大(〉35%),随着重建内核变得锐利,测量误差近似呈线性递减关系,分别在B60f和B50f达到最小(〈5%),PC的测量误差范围为6.93Voo~14.47%。结论测量壁厚约为1~2mm的气道体模时,FWHM和ZC方法在B50f时误差最小;而PC方法受重建内核影响小,测量结果稳定。Objective To assess the accuracy of thoracic airway measurement with different methods on CT images ob- tained under different conditions. Methods The airway phantom, which was made from poly-acryl tubes and foam, was scanned with Siemens 64-multi-detector row CT scanner. Data obtained with different FOV were reconstructed into CT im- ages with different reconstruction kernels. Then CT images of airway phantom were measured by using full width at half maximum (FWHM), zero crossing (ZC) and phase congruency (PC), and the measurement errors were compared. Results For tubes with the thickness of about 1.5--2.0 mm, when reconstruction kernels were varied from B30f to B60f, the measurement errors by FWHM and ZC were varied from positive to increased negative and reached the minimum at B45f (~11~), and those by PC were relatively steady. For tubes with thickness of about 1 ram, the measurement errors by FWHM and ZC reached maximum (~ 35 ~//0 ) at B30f, then dropped near linearly with the reconstruction kernel becoming sharper, and reached the minimum at B60f and BS0f (~5%). The range of measurement errors by PC was 6.93%-- 14.47%. Conclusion To measure airway phantoms with thickness of 1--2 mm, the total error by FWHM and PC is mini- mum at B50f, while PC is less affected by reconstruction kernels and remains steady.
关 键 词:体层摄影术 X线计算机 重建内核 视野 气道体模
分 类 号:R814.42[医药卫生—影像医学与核医学]
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.28