摘 要:植入式心律转复除颤器(implantable cardioverter defibrillator,ICD)的问世及发展改变了心源性猝死患者的命运,随着ICD技术改进及程控参数的多元化,其应用更为普及。ICD是治疗心血管疾病的利器,已成为心脏科医生防治恶性心律失常的重要手段。本文回顾了近5年来植入的42例ICD患者临床资料,旨在探讨ICD治疗室性心律失常的效果。1对象与方法 1.1对象自2008年7月以来。Since 2008,42 patients were implanted ICD in our hospital.Of them,34patients were followed up(20male,62.5%),including 9 single chamber ICD,1 double chamber ICD and 25CRT-D(one replaced). There were 1 726 ventricular tachycardia(VT)episodes detected(510VT-treated and 1 216VT-monitored episodes),moreover,99fast ventricular tachycardia(FVT)and 33ventricular fibrillator(VF).The 589VT/FVT were treated by anti-tachycardia pacing(ATP),among them,387(65.7%)were successfully stopped and 12episodes(2%)were got worse after ATP.The 18VT/FVT episodes were treated by low-energy converting and 8(44.4%)were stopped.Of VF episodes,3imappropriate detection were absorted after charging,15episodes were imappropriate discharged(3patients,8.8%),1episode were stopped by ATP,17episodes were treated by high-energy defibrillate and 15(88.2%)were successfully stopped.
关 键 词:植入式心律转复除颤器 室性心律失常 电击 抗心动过速起搏