作 者:富青[1] 余建明[1] 孔祥闯[1] 刘永华[1] 雷子乔[1] 韩萍[1] 孔祥泉[1] 梅豪
机构地区:[1]华中科技大学同济医学院附属协和医院放射科,武汉430022
出 处:《中华放射医学与防护杂志》2014年第1期62-66,共5页Chinese Journal of Radiological Medicine and Protection
摘 要:目的探讨80kV低管电压对128层螺旋CT肺动脉成像(CTPA)图像质量及辐射剂量的影响。方法对临床怀疑肺栓塞患者60例行128层螺旋CT肺动脉成像检查,并采用随机数字表法分为低管电压80kV组和常规管电压120kV组各30例,两组均开启自动管电流调制技术。记录两组的CT容积剂量指数(CTDIvol)、剂量长度乘积(DLP),计算有效剂量(E)值;测量图像背景噪声(BN)、肺动脉强化CT值(sI),计算信噪比(SNR)、对比噪声比(CNR)。比较两组的有效剂量E、SNR、CNR以及图像质量。结果80kV组的E为(0.99±0.27)mSv,显著低于120kV组的(3.02±0.87)mSv(t=12.281,P〈0.05)。80kV组SI、BN均显著高于120kV组(t=-3.377、-5.855,P〈0.05),SNR、CNR和图像质量评分差异均无统计学意义(P〉0.05)。结论与常规管电压120kV扫描方案相比,采用低管电压80kV,结合自动管电流调制技术,可以在保证图像质量的同时,有效降低辐射剂量,在多层螺旋CT肺动脉成像中具有较高的临床应用价值。Objective To investigate the influence of 80 kV on dose reduction and image quality in 128-slice CT pulmonary angiography (CTPA) compared with the conventional 120 kV. Methods Sixty patients suspected of pulmonary embolism (PE) who underwent CTPA were randomly divided into 80 kV and 120 kV groups by random digits table method. All the patients underwent CTPA with automatic tube current modulation. The values of volume CT dose index ( CTDIvo1) , dose length product (DLP) were recorded and effective dose (E) was calculated. The attenuation was measured in central and peripheral pulmonary arteries as well as background noise (BN) , then signal-to-noise-ratio (SNR) and contrast-to- noise-ratio (CNR) were calculated. The radiation exposure, pulmonary arterial enhancement, background noise, SNR, CNR and scores of image quality by a 5-point scale were compared between two groups. Results E value was significantly lower at 80 kV [ (0.99 ± 0.27 ) roSy ] compared with 120 kV [ (3.02 ± 0.87) mSv, t = 12. 281, P 〈 0. 05 ]. The mean attenuation value of pulmonary arteries as well as BN in 80 kV group were significantly higher than in 120 kV group ( P 〈 0.05 ). SNR and CNR did not differ significantly between two groups. No significant difference was detected on scores of image quality (P 〉 0.05). Conclusions The application of 80 kV protocol combined with automatic tube current modulation scanning in 128-slice MSCT pulmonary angiography could significantly reduce radiation dose compared with the conventional 120 kV protocol without deterioration in diagnostic image quality.
关 键 词:CT肺动脉成像 辐射剂量 管电压 X射线计算机体层摄影
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.222