血细胞分析仪自动全血和预稀释模式的结果分析  被引量:2

Analysis on the test results of hematology analyzers in the automatic whole-blood mode and in the pre-dilution mode

在线阅读下载全文

作  者:张宪华[1] 

机构地区:[1]河南省安阳市第五人民医院检验科,河南安阳455000

出  处:《中国现代医生》2014年第6期74-76,共3页China Modern Doctor

摘  要:目的分析血细胞分析仪自动全血和预稀释模式的结果。方法在迈瑞BC-5600全自动血细胞分析仪上对全血质控物用自动全血和预稀释两种模式同时各行30次检测并进行统计学分析。结果自动全血和预稀释模式白细胞(WBC)、红细胞(RBC)、血红蛋白(HGB)、血细胞比容(HCT)、血小板(PLT)的结果和它们的变异系数(CV)分别为(7.3±0.07,0.97;4.97±0.03,0.60;130.0±0.91,0.70;41.2±0.26,0.63;228.4±3.23,1.41和8.0±0.35,4.38;4.13±0.19,4.60;124.2±5.98,4.81;38.6±1.80,4.66;230.1±9.21,4.00。对应项目的 t值和P值分别为10.74,<0.01;7.57,<0.01;5.43,<0.01;7.83,<0.01;0.95,>0.05。结论采用自动全血模式检测结果比预稀释模式的变异系数小,精密度高,重复性好。Objective To analyze the test results of hematology analyzers in the automatic whole-blood mode and in the pre-dilution mode. Methods Mindray BC-5600 automatic hematology analyzer was applied for test of whole blood quality control materials for 30 times in the automatic whole-blood mode and the pre-dilution mode. Statistical analysis was conducted in the study. Results The test results and variation coefficients of white blood cell, red blood cell, hemoglobin, hematocrit and platelet in the automatic whole-blood mode were 7.3±0.07, 0.97; 4.97±0.03, 0.60; 130.0± 0.91, 0.70; 41.2±0.26, 0.63; 228.4±3.23, 1.41 respectively; and the test results and variation coefficients in the pre-dilution mode were 8.0±0.35, 4.38; 4.13±0.19, 4.60; 124.2±5.98, 4.81; 38.6±1.80, 4.66; 230.1±9.21, 4.00 respectively. The t value and P value for the five indicators were 10.74, 〈0.01; 7.57, 〈0.01; 5.43, 〈0.01; 7.83, 〈0.01; 0.95, 〉0.05 respectively. Conclusion Compared with the pre-dilution mode, the automatic whole-blood mode has smaller variation coefficient, higher precision and better repeatability.

关 键 词:血细胞分析仪 自动全血模式 预稀释模式 

分 类 号:R446.1[医药卫生—诊断学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象