机构地区:[1]南京医科大学第二附属医院心内科,江苏南京210011
出 处:《中国医疗设备》2014年第4期120-122,共3页China Medical Devices
摘 要:目的探讨无创血压监护仪在心房纤颤病人血压测量中的准确性。方法对60例心房纤颤病人分别采用无创血压监护仪和汞柱台式血压计进行血压监测,对比其测量结果。结果无创血压监护仪和汞柱台式血压计血压测量结果比较,收缩压无显著差异(P>0.05),而舒张压有显著差异(P<0.05)。结论无创血压监护仪在心房纤颤病人血压测量中的准确性较差,值得临床注意。Objective To investigate the accuracy of non-invasive blood pressure monitor in blood pressure measurement of patients with atrial fibrillation. Methods The blood pressure of 60 patients with atrial fibrillation was measured by using non-invasive blood pressure monitor and mercury sphygmomanometer, respectively. Then the measuring results of blood pressure of non-invasive blood pressure monitor were compared with those of mercury sphygmomanometer. Results There was no signiifcant difference between the measuring results of blood pressure of non-invasive blood pressure and those of mercury sphygmomanometer on systolic pressure (P〉0.05) while there were signiifcant differences between the measuring results of blood pressure of non-invasive blood pressure and those of mercury sphygmomanometer on diastolic pressure (P〈0.05). Conclusion It is inaccurate to measure the blood pressure of patients with atrial ifbrillation by using non-invasive blood pressure monitor, which should be paid attention to in clinic.