输液泵质量检测技术规范中测试时间选择问题的探讨  被引量:4

Discussion on Problems of Testing Time Selection in Technology Standard for Quality Control of Infusion Pump

在线阅读下载全文

作  者:许耀良[1] 于洋[1] 陈基明[1] 

机构地区:[1]中国人民解放军空军总医院医学工程科,北京100142

出  处:《中国医疗设备》2014年第5期55-57,共3页China Medical Devices

摘  要:由于输液泵的工作原理及特点,在质量控制检测过程中的流量检测,特别是超范围流量检测耗费的时间较长。在保证检测准确性的前提下,我们对检测工作进行了一系列实验。通过对实验数据进行分析,对规范中有关超标输液泵的延时检测方法进行了调整,从而减少了大量无效工作时间,提高了检测效率。Due to the working principle and characteristics of infusion pump, fl ow detection, especially the fl ow detection which is out of the scope of technology standard would take longer to be fi nished in the detection process of infusion pump quality control. Based on the original detection accuracy, a series of experiments were conducted during the detection process of infusion pump quality control. The detection methods of unqualifi ed infusion pump were adjusted by analyzing the experimental data in practical work. Thus the amount of ineffective work time have been reduced and the detection effi ciency of infusion pump have been improved.

关 键 词:输液泵 时间检测 流量检测 质量控制 

分 类 号:TH789[机械工程—仪器科学与技术]

 