降低管电流对CT结肠仿真内镜图像质量及辐射剂量的影响  被引量:2

THE IMPACT OF LOWER TUBE CURRENT COLONIC GEM ENERGY SPECTRUM CT SCANNING ON IMAGE QUALITY AND RADIATION DOSAGE

在线阅读下载全文

作  者:李家言[1] 袁捷[1] 韦兰珍 廖海[1] 黄增超[1] 刘熙荣[1] 袁文昭[1] 何欣[1] 张锡流[1] 黄乔统[1] 

机构地区:[1]广西中医药大学第一附属医院放射科,南宁530023

出  处:《广西医科大学学报》2014年第2期232-235,共4页Journal of Guangxi Medical University

基  金:广西壮族自治区卫生厅科研自筹资助项目(No.Z2013168)

摘  要:目的:探讨降低管电流宝石能谱CT低剂量扫描对结肠仿真内镜成像图像质量及辐射剂量的影响。方法:对45例临床怀疑为结肠息肉病变的患者平均分为3组,行宝石能谱CT降低管电流结肠仿真内镜成像,其中按固定管电流分别为50,30,15mAs进行扫描,其他参数按常规进行,并对3组图像质量主观评分及有效辐射剂量进行评价。结果:固定管电流为50mAs(A组)时,结肠仿真内镜成像(CTVC)及透明重组图像(Raysum)平均评分为(2.93±0.26)分,平均容积CT剂量指数(CTDIvol)与剂量长度乘积(DLP)分别为(1.68±0.00)mGy和(80.62±3.44)mGy-cm;30mAs(B组)时,上述平均评分为(2.53±0.52)分,剂量平均分别为(1.01±0.01)mGy和(45.09±5.54)mGy-cm;15mAs(C组)时,上述平均评分为(1.93±0.70)分,剂量平均分别为(0.05±0.00)mGy和(22.16±1.96)mGy-cm。3组CTDIvol与DLP平均分别降低了80.05%、88.00%、94.06%和80.73%、88.95%、94.57%。结论:降低管电流在CTVC成像中,在管电流与辐射剂量分别降低88%以上的情况下,仍能兼顾图像质量,而降低至94.57%时,对图像质量影响较大。Objective:To explore the role of lower tube current gem energy spectrum low dose CT scanning for colon virtual endoscopy imaging scan image quality and dose effect.Methods:Colon virtual endoscopy imaging in 45 patients with suspected lesions of colon polyps was performed with lower tube current gem energy spectrum CT scanning,and the patients were evenly divided into 3 groups according to the fixed tube current of 50,30 and 15 mAs.Other parameters were performed according to routine.The subjective grading and effective radiation dose of the image quality were evaluated.Results:When the tube current was fixed as 50 mAs (group A),CTVC and the raysum image average scoring was 2.93±0.26 points,average CTDIvol and DLP was 1.68 mGy and 80.62 mGy cm; respectively.When the tube current was fixed as 30 mA (group B),above average scoring was 2.53±0.52 points,dose average was 1.01 mGy and 45.09 mGy cm; respectively.When the tube current was fixed as 15 mA (group C),above average scoring was 1.93±0.70 points,dose average was 0.5 mGy and 22.16 mGy cm; respectively.The average of CTDIvol andDLPin three groups decreased 80.05%,88% and 94.06%; and 80.73%,88.95% and 94.57%; respectively.Conclusion:The lower tube current CTVC imaging,in the case of radiation dose to reduce more than 88%,is still able to combine image quality,but reduced to 94%,the picture quality influence is big.

关 键 词:结肠成像 体层摄影术 X线计算机 辐射剂量 

分 类 号:R81[医药卫生—放射医学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象