摘 要:目的:评价BECKMAN ACCESS2全自动微粒子化学发光免疫分析系统相关性能。方法:以测定游离三碘甲状腺原氨酸(FT3)为例,收集患者高、中、低值新鲜血清混合标本,应用CLSI颁布的EP10-A2对ACCESS2全自动化学发光免疫分析系统进行初步评价。结果:通过CLSI EP10-A2程序评价显示ACCESS2检测系统偏差、不精密度、交叉污染、线性漂移等因素都在可接受范围内,对精密度无明显影响。结论:ACCESS2全自动发光免疫分析系统分析性能可靠,适合临床实验室使用。Objective:To evaluate the performance of Beckman Access2 automated chemiluminescent immunoassay system in order to know its status. Method:Using the CLSI document EP10-2A for preliminary evaluation methods to approve the high,medium and low fresh serum of patients which were adopted as samples and was assayed FT3 in Beckman Access2 testing system.Result:The factors of carryover and drift,system deviation,imprecision,cross contamination in the ballpark,they had no effect on precision.Conclusion:The performance of Beckman Access2 automated chemiluminescent system is reliable and suitable for clinical laboratory to use.