FibroTouch及FibroScan~实施受控衰减参数评价肝细胞脂肪变与肝脏病理比较分析  被引量:10

Comparative analysis of controlled attenuation parameters in evaluation of liver steatosis using FibroTouch and FibroScan~

在线阅读下载全文

作  者:朱梦飞[1] 刘静[1] 王洁[1] 陈公英[1] 娄国强[1] 施军平[1] 

机构地区:[1]杭州师范大学附属医院,杭州310015

出  处:《中国实用内科杂志》2014年第6期589-593,共5页Chinese Journal of Practical Internal Medicine

基  金:杭州市科技计划项目(20110833B07);杭州市卫生局项目(2011Z006)

摘  要:目的评价瞬时弹性超声诊断仪FibroTouch、FibroScan实施受控衰减参数(CAP)无创定量诊断脂肪肝与肝脏穿刺病理结果的对比及相关分析。方法对2013年7—10月在杭州师范大学附属医院对病理确诊的非酒精性脂肪性肝病(NAFLD)患者及慢性乙型肝炎患者(CHB)应用FibroTouch、FibroScan-502机型测定肝脏脂肪含量(CAP值)。结果41例CHB患者与20例NAFLD患者同时接受FibroTouch、FibroScan检查,测定的CAP与患者体质量指数(BMI,r分别为0.42,0.61;P分别为0.02,0.000)及肝细胞脂肪变呈正相关(r分别为0.54,0.56;P均〈0.001);33例病理诊断肝细胞脂肪变〉5%的患者FibroTouch与FibroScan测定的CAP值均显著高于28例无脂肪变的CAP值{FibroTouch:(252.18±41.23)dB/m比(220.68±54.75)dB/m,P=0.04;FibroScan:(291.61±56.80)dB/m比(215.75±45.11)dB/m,P=0.000];FibroScan测定肝细胞脂肪变〈30%患者的CAP值、肝细胞脂肪变30%~60%患者的CAP值、〉60%脂肪变组的CAP值间存在差异(F=6.82,P=0.004),且肝细胞脂肪变〈30%患者的CAP值低于〉60%脂肪变组的CAP值[(258.73±52.54)dB/m比(327.42±49.08)dB/m,P=0.04]。但FibroTouch对上述3组患者测定的CAP值间差畀无统计学意义(F=2.30,P=0.12)。结论FibroTouch与Fi—broScan测定的CAP值均能评价肝细胞脂肪变,CAP值的高低与患者的BMI大小、肝病理脂肪变严重程度呈正相关。FibroScan测定的CAP值能区分轻度与重度的脂肪变,而FibroTouch测定的CAP值不能区分,可能与本组实验病例过少有关,有待进一步研究。Objective To compare the controlled attenuation parameters (CAP) measurement of liver steatosis using FibroTouch and FibroScan. Methods Patients with non-alcohol fatty liver disease (NAFLD) and chronic hepatitis B (CHB) were included from July to October 2013 and liver biopsy was completed on these patients. They were examined to collect the CAP values by FibroTouch and FibroSean-502 respectively. Results A total of 41 patients with CHB and 20 patients with NAFLD were recruited. CAP values of FibroScan and FibroTouch are positively correlated with BMI (r = 0.42,0. 61; P = 0. 02,0. 000) and degree of liver steatosis (r = 0. 54,0. 56;P = 0. 001,0. 001 ). CAP values of 33 patients with 〉 5% fat degeneration of liver cells were significantly higher than those of patients with no fat degeneration (FibroTouch [252. 18 ±41.23]dB/m vs [220.68 ±54.75]dB/m,P =0.04;FibroScan[291.61 ±56.80] B/m vs [215.75± 45.11 ] dB/m ,P = 0. 000). CAP values of patients with liver steatosis 〈 30% were significantly lower than those of patients with liver steatosis 〉 60% by FibroScan ( [ 258.73 -+ 52. 54 ] dB/m vs [ 327.42 _+ 49. 08 ] dB/m,P = 0. 04). But there were no differences in CAP values among patients with liver steatosis less than 30% ,between 30% -60% ,and more than 60%. Conclusion CAP values by FibroTouch and FibroScan are positively correlated with body mass and degree of liver steatosis. CAP value is a good and noninvasive index to evaluate degree of liver steatosis.

关 键 词:瞬时弹性超声诊断仪 受控衰减参数 病理 肝细胞脂肪变 

分 类 号:R445.1[医药卫生—影像医学与核医学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象