除颤器释放能量溯源方法的研究  被引量:14

Study on Traceability Method of Defibrillator Released Energy

在线阅读下载全文

作  者:贾建革[1] 张亚冬[2] 武文君[1] 李岩峰[1] 段新安[1] 赵正男[1] 邵海明[3] 

机构地区:[1]中国人民解放军总后勤部卫生部药品仪器检验所,北京100071 [2]第四军医大学 生物医学工程学院,陕西西安710032 [3]中国计量科学研究院,北京100013

出  处:《中国医疗设备》2014年第6期54-57,共4页China Medical Devices

基  金:国家科技支撑计划课题项目(2011BAI02B04);军事医学计量科研专项课题(2011-JL2-062)

摘  要:本文介绍了除颤器质量控制现状,分析了对除颤器分析仪进行计量校准存在的问题。然后从理论上提出了一种用来完善除颤器分析仪计量校准的标准器具即除颤能量标准源的研制方法。最后根据医学计量体系的要求,研究并提出了一套完整的除颤器释放能量参数的溯源体系。This paper describes the quality control status of defibrillator and analyzes the problems exist in the metrology calibration process of defibrillator analyzer. Then the theoretical development method of a new standard apparatus called standard defibrillator energy source whose design goal is to improve the metrology calibration process of defibrillator analyzer is proposed. A complete traceability system of defibrillator released energy is also proposed according to the requirements of medical metrology system in this paper.

关 键 词:除颤器 除颤能量 质量控制 医学计量 除颤分析仪 

分 类 号:R197.39[医药卫生—卫生事业管理] TH77[医药卫生—公共卫生与预防医学]

 