品管圈活动在过氧化氢低温等离子灭菌质量控制中的应用  被引量:8

Application of Quality Control Circle Program on Quality Control of Hydrogen Peroxide Low Temperature Plasma

在线阅读下载全文

作  者:刘肖[1] 

机构地区:[1]广西百色市人民医院,广西百色533000

出  处:《中国医学创新》2014年第17期113-115,共3页Medical Innovation of China

基  金:广西百色市科技计划项目(20111601)

摘  要:目的:探讨品管圈活动在过氧化氢低温等离子灭菌质量控制中的应用效果。方法:成立品管圈小组,对过氧化氢低温等离子灭菌循环程序中断事件进行原因分析,制定相应对策,对关键环节实施质量控制及持续改进。结果:品管圈活动开展后,灭菌循环程序中断率由3.8%下降为0.87%,显著提高了灭菌循环成功率。结论:应用品管圈管理工具,可做好过氧化氢低温等离子灭菌质量的环节控制。Objective: To explore the application effect of quality control circle program on quality control of hydrogen peroxide low temperature plasma.Method: We set up a team of quality control circle, analyzed the reasons of interrupt events in the application of hydrogen peroxide low temperature plasma sterilization cycle.Then we established corresponding countermeasures and kept continuous quality monitoring to the key points.Result: After quality control circle activities, the rate of interrupt events in sterilization cycle decreased from 3.8% to 0.87%, the success rate of sterilization cycle significantly raised.Conclusion: Application of quality control circle program is favorable in the quality control of hydrogen peroxide low temperature plasma.

关 键 词:供应室 品管圈 过氧化氢低温等离子灭菌 质量控制 

分 类 号:R187[医药卫生—流行病学]

 